ORTHOPEDIATRICS CANADA ULC FDA Inspection Record

FEI 3000327445 · Inspections, Form 483s, warning letters and import refusals

ORTHOPEDIATRICS CANADA ULC is an FDA-registered facility (FEI 3000327445) located in Laval Quebec, CA. RegDataLab tracks 3 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Mar 1, 2018: Voluntary Action Indicated (VAI)
Inspections on record
3 (2 OAI, 1 VAI)
Last OAI inspection
Jul 25, 2013
Warning letters
1 listed by FDA for this FEI, latest Oct 26, 2010
Recalls as the recalling firm
10, latest Jul 4, 2024
Import refusals
2

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 1, 2018Voluntary Action Indicated (VAI)Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Production and Process Change Procedures, lack of or Inad. (21 CFR 820.70(b))
and 1 more
Jul 25, 2013Official Action Indicated (OAI)Procedures not adequately established or maintained (21 CFR 820.40)
Lack of or inadequate procedures (21 CFR 820.100(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Justification for not reporting (21 CFR 806.20(b)(4))
Mar 31, 2010Official Action Indicated (OAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 4, 2024Class IIZ-2941-2024FRT250 - 4.0-6.4 CARTRIDGE. Component of the Fassier-Duval Telescopic IM System
During the assembly of the FRT250 cartridge assembly, an incorrect component was used. Specifically, the FRT240-HEX (2) was used instead of…
Apr 3, 2020Class IIIZ-2038-2020Orthopedic manual Surgical Instrument handle for Simple Locking Intra Medullary (SLIM)…
The lot number on the pediatric orthopedic implant driver instrument handle does not match the lot number on the label.
Oct 2, 2019Class IIZ-0270-2020SCFE (slipped capital femoral epiphysis) Knobs for Driver, 6.5mm and 7.3 mm Orthopedic…
There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.
May 29, 2019Class IIZ-2278-2019SLIM Handle - catalogue #: SLM-HND100 a component of Simple Locking Intramedullary (SLIM)…
Handle could jam with the Driver due to the detachment of an internal Connector Ring.
Oct 20, 2015Class IIZ-0324-2016SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free…
SCFE Driver's threaded shaft broke close to the knob when a manual force was applied during the fixation of the screw.
Jun 15, 2015Class IIZ-1848-2015Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System…
SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high…
Apr 10, 2015Class IIZ-1546-2015Female Driver, Part number FDr101, instrument used for the insertion of the Fassier-Duval…
The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may…
Feb 20, 2015Class IIZ-1362-2015GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number…
Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures…
Mar 5, 2014Class IIZ-1472-2014Fassier-Duval IM Telescopic System. Orthopedic intramedullary rod implant. Catalogue…
The outer package label contained the incorrect length of the device. There is a label on the outer package and an additional label on the…
Feb 7, 2014Class IIZ-1448-2014Pega Medical Inc., Fassier-Duval IM Telescopic System, male component (solid shaft), Part…
The Male Components in this lot are made of material with lower strength than manufacturing specifications.

FDA record score

6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 15, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

13 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2941-2024, this FEI as the recalling firmJul 4, 20244 x 0.62.4
Inspection classified VAIMar 1, 20186 x 0.251.5
Class II recall Z-0270-2020, this FEI as the recalling firmOct 2, 20194 x 0.251
Class II recall Z-2278-2019, this FEI as the recalling firmMay 29, 20194 x 0.251
Class II recall Z-0324-2016, this FEI as the recalling firmOct 20, 201510 years or older, so it adds no points0
Class II recall Z-1848-2015, this FEI as the recalling firmJun 15, 201510 years or older, so it adds no points0
Class II recall Z-1546-2015, this FEI as the recalling firmApr 10, 201510 years or older, so it adds no points0
Class II recall Z-1362-2015, this FEI as the recalling firmFeb 20, 201510 years or older, so it adds no points0
Class II recall Z-1472-2014, this FEI as the recalling firmMar 5, 201410 years or older, so it adds no points0
Class II recall Z-1448-2014, this FEI as the recalling firmFeb 7, 201410 years or older, so it adds no points0
Inspection classified OAIJul 25, 201310 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsOct 26, 201010 years or older, so it adds no points0
Inspection classified OAIMar 31, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 15, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 15, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 14, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 15, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing ORTHOPEDIATRICS CANADA ULC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check ORTHOPEDIATRICS CANADA ULC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
ORTHOPEDIATRICS CANADA ULC
Location
Laval Quebec (CA)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.