ENDOPLUS FDA Inspection Record
ENDOPLUS is an FDA-registered facility (FEI 3000719482) located in Mundelein, IL, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 25, 2026: No Action Indicated (NAI)
- Inspections on record
- 6 (1 VAI, 5 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Jul 15, 2013
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 25, 2026 | No Action Indicated (NAI) | No citations listed |
| Sep 8, 2022 | No Action Indicated (NAI) | No citations listed |
| Aug 2, 2017 | No Action Indicated (NAI) | No citations listed |
| May 19, 2015 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Documentation (21 CFR 820.100(b)) |
| Feb 26, 2013 | No Action Indicated (NAI) | Lack of Written MDR Procedures (21 CFR 803.17) |
| Jun 19, 2009 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 15, 2013 | Class II | Z-2264-2013 | ENDOPLUS; Precision Laparoscopic Instruments; REF 346VA; Description 5mm Atraumatic… and 1 more product Multiple complaints reporting that the moveable jaw has broken during laparoscopic procedures. |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
2 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2264-2013, this FEI as the recalling firm | Jul 15, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 19, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing ENDOPLUS as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ENDOPLUS yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Endoplus
- Location
- Mundelein, IL (US)
- Source
- device