What FDA found at this site
- Latest inspection
- Oct 23, 2025: No Action Indicated (NAI)
- Inspections on record
- 4 (1 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 4, latest Aug 20, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 23, 2025 | No Action Indicated (NAI) | No citations listed |
| Aug 2, 2019 | No Action Indicated (NAI) | No citations listed |
| Jan 13, 2016 | No Action Indicated (NAI) | No citations listed |
| Mar 18, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 20, 2026 | Class II | Z-0192-2027 | Think Surgical Square Driver
REF: 109126 Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin… |
| May 8, 2026 | Class II | Z-2716-2026 | TMINI Femur Tracking Array REF 109003
UDI-DI code: M9461090030
TMINI Tibia Tracking… Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array. |
| Feb 12, 2024 | Class II | Z-1318-2024 | Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with… There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient… |
| Mar 30, 2018 | Class II | Z-2605-2018 | Think Surgical Cutter, Flat, ¿6.2mm x 144mm. For use only with the TCAT Computer-Assisted… There is a potential for the cutter head to detach from the cutter shaft. |
FDA record score
11 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0192-2027, this FEI as the recalling firm | Aug 20, 2026 | 4 x 1 | 4 |
| Class II recall Z-2716-2026, this FEI as the recalling firm | May 8, 2026 | 4 x 1 | 4 |
| Class II recall Z-1318-2024, this FEI as the recalling firm | Feb 12, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-2605-2018, this FEI as the recalling firm | Mar 30, 2018 | 4 x 0.25 | 1 |
| Inspection classified VAI | Mar 18, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |