TRIMED, INC. FDA Inspection Record

FEI 3001236812 · Inspections, Form 483s, warning letters and import refusals

TRIMED, INC. is an FDA-registered facility (FEI 3001236812) located in Santa Clarita, CA, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Jun 3, 2024: Voluntary Action Indicated (VAI)
Inspections on record
6 (1 OAI, 4 VAI, 1 NAI)
Last OAI inspection
Mar 18, 2016
Warning letters
1 listed by FDA for this FEI, latest Jun 30, 2016
Recalls as the recalling firm
11, latest Mar 19, 2024

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jun 3, 2024Voluntary Action Indicated (VAI)Design verification - documentation (21 CFR 820.30(f))
Documentation (21 CFR 820.100(b))
Aug 14, 2019No Action Indicated (NAI)No citations listed
Nov 17, 2017Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design control - no procedures (21 CFR 820.30(a))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Documentation (21 CFR 820.80(e))
and 2 more
Jan 12, 2017Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design control - no procedures (21 CFR 820.30(a))
Software validation for automated processes (21 CFR 820.70(i))
Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
and 7 more
Mar 18, 2016Official Action Indicated (OAI)Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Design control - no procedures (21 CFR 820.30(a))
Software validation for automated processes (21 CFR 820.70(i))
Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
and 9 more
Jan 25, 2012Voluntary Action Indicated (VAI)Design history file (21 CFR 820.30(j))
Investigation of device failures (21 CFR 820.198(c))
Equipment control activity documentation (21 CFR 820.72(a))
Quality Audits - defined intervals (21 CFR 820.22)
and 1 more

Recalls

Showing the 10 most recent of 11 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Mar 19, 2024Class IIZ-1791-2024TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx… and 1 more product
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
Nov 28, 2023Class IIZ-0932-2024Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S…
Plates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes.
Dec 22, 2022Class IIZ-1101-2023RipCord Syndesmosis Button, REF: STA001K
Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia…
Sep 28, 2018Class IIZ-0355-2020TriMed Countersink 1.7mm, HCS, REF HSINK-1.7, nonsterile, Rx. The firm name on the label… and 4 more products
The epoxy on Apex manufactured epoxy-coated instruments can pop off.
Aug 24, 2018Class IIZ-0118-2020TriMed Dorsal Wrist Hook Plate, Dorsal, 4-Hole, REF WHD-4, contained in a tray, insert, or…
Modifications to the specification were not appropriately qualified through the company's design control process which may result in…
Aug 24, 2018Class IIZ-1214-2019TriMed Hex Cortical Screw 4.0, 40 mm, REF HEX4.0-40, non-Sterile, RX, a component of the…
The bone screw is shorter than the intended design specification.
Aug 15, 2018Class IIZ-0366-2020TriMed Olecranon Hook Plate, Lt, 6-Holes, REF: OHOOKL-6, Non-Sterile, RX only, (01)…
The firm has become aware that the angled-hole of certain semi-tubular design bone plates may have been modified from its specification, and…
Apr 16, 2018Class IIZ-2877-2018Ulnar Osteotomy Compression Plate, Short Model: UOCP-S Product Usage: The UOCP-S is…
Two instrument holes located on the side of the plate, which act to align the saw guide are located on the wrong side of the plate
Apr 13, 2018Class IIZ-2262-2018Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b)… and 3 more products
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required…
Jan 11, 2018Class IIZ-1373-2018Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver…
Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate…

FDA record score

21 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

17 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIJun 3, 20246 x 0.63.6
Class II recall Z-1791-2024, this FEI as the recalling firmMar 19, 20244 x 0.62.4
Class II recall Z-0932-2024, this FEI as the recalling firmNov 28, 20234 x 0.62.4
Class II recall Z-1101-2023, this FEI as the recalling firmDec 22, 20224 x 0.62.4
Inspection classified VAINov 17, 20176 x 0.251.5
Inspection classified VAIJan 12, 20176 x 0.251.5
Class II recall Z-0355-2020, this FEI as the recalling firmSep 28, 20184 x 0.251
Class II recall Z-0118-2020, this FEI as the recalling firmAug 24, 20184 x 0.251
Class II recall Z-1214-2019, this FEI as the recalling firmAug 24, 20184 x 0.251
Class II recall Z-0366-2020, this FEI as the recalling firmAug 15, 20184 x 0.251
Class II recall Z-2877-2018, this FEI as the recalling firmApr 16, 20184 x 0.251
Class II recall Z-2262-2018, this FEI as the recalling firmApr 13, 20184 x 0.251
Class II recall Z-1373-2018, this FEI as the recalling firmJan 11, 20184 x 0.251
Class II recall Z-2849-2016, this FEI as the recalling firmAug 29, 201610 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJun 30, 201610 years or older, so it adds no points0
Inspection classified OAIMar 18, 201610 years or older, so it adds no points0
Inspection classified VAIJan 25, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing TRIMED, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check TRIMED, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
TriMed Inc.
Location
Santa Clarita, CA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.