TRIMED, INC. FDA Inspection Record
TRIMED, INC. is an FDA-registered facility (FEI 3001236812) located in Santa Clarita, CA, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jun 3, 2024: Voluntary Action Indicated (VAI)
- Inspections on record
- 6 (1 OAI, 4 VAI, 1 NAI)
- Last OAI inspection
- Mar 18, 2016
- Warning letters
- 1 listed by FDA for this FEI, latest Jun 30, 2016
- Recalls as the recalling firm
- 11, latest Mar 19, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 3, 2024 | Voluntary Action Indicated (VAI) | Design verification - documentation (21 CFR 820.30(f)) Documentation (21 CFR 820.100(b)) |
| Aug 14, 2019 | No Action Indicated (NAI) | No citations listed |
| Nov 17, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design control - no procedures (21 CFR 820.30(a)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Documentation (21 CFR 820.80(e)) and 2 more |
| Jan 12, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design control - no procedures (21 CFR 820.30(a)) Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) and 7 more |
| Mar 18, 2016 | Official Action Indicated (OAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design control - no procedures (21 CFR 820.30(a)) Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) and 9 more |
| Jan 25, 2012 | Voluntary Action Indicated (VAI) | Design history file (21 CFR 820.30(j)) Investigation of device failures (21 CFR 820.198(c)) Equipment control activity documentation (21 CFR 820.72(a)) Quality Audits - defined intervals (21 CFR 820.22) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Mar 19, 2024 | Class II | Z-1791-2024 | TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx… and 1 more product Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque. |
| Nov 28, 2023 | Class II | Z-0932-2024 | Volar Bearing Plate.
REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S… Plates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes. |
| Dec 22, 2022 | Class II | Z-1101-2023 | RipCord Syndesmosis Button, REF: STA001K Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia… |
| Sep 28, 2018 | Class II | Z-0355-2020 | TriMed Countersink 1.7mm, HCS, REF HSINK-1.7, nonsterile, Rx. The firm name on the label… and 4 more products The epoxy on Apex manufactured epoxy-coated instruments can pop off. |
| Aug 24, 2018 | Class II | Z-0118-2020 | TriMed Dorsal Wrist Hook Plate, Dorsal, 4-Hole, REF WHD-4, contained in a tray, insert, or… Modifications to the specification were not appropriately qualified through the company's design control process which may result in… |
| Aug 24, 2018 | Class II | Z-1214-2019 | TriMed Hex Cortical Screw 4.0, 40 mm, REF HEX4.0-40, non-Sterile, RX, a component of the… The bone screw is shorter than the intended design specification. |
| Aug 15, 2018 | Class II | Z-0366-2020 | TriMed Olecranon Hook Plate, Lt, 6-Holes, REF: OHOOKL-6, Non-Sterile, RX only, (01)… The firm has become aware that the angled-hole of certain semi-tubular design bone plates may have been modified from its specification, and… |
| Apr 16, 2018 | Class II | Z-2877-2018 | Ulnar Osteotomy Compression Plate, Short
Model: UOCP-S
Product Usage:
The UOCP-S is… Two instrument holes located on the side of the plate, which act to align the saw guide are located on the wrong side of the plate |
| Apr 13, 2018 | Class II | Z-2262-2018 | Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b)… and 3 more products During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required… |
| Jan 11, 2018 | Class II | Z-1373-2018 | Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver… Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate… |
FDA record score
21 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
17 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jun 3, 2024 | 6 x 0.6 | 3.6 |
| Class II recall Z-1791-2024, this FEI as the recalling firm | Mar 19, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0932-2024, this FEI as the recalling firm | Nov 28, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1101-2023, this FEI as the recalling firm | Dec 22, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Nov 17, 2017 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Jan 12, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0355-2020, this FEI as the recalling firm | Sep 28, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0118-2020, this FEI as the recalling firm | Aug 24, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1214-2019, this FEI as the recalling firm | Aug 24, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0366-2020, this FEI as the recalling firm | Aug 15, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2877-2018, this FEI as the recalling firm | Apr 16, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2262-2018, this FEI as the recalling firm | Apr 13, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1373-2018, this FEI as the recalling firm | Jan 11, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2849-2016, this FEI as the recalling firm | Aug 29, 2016 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jun 30, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Mar 18, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 25, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing TRIMED, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check TRIMED, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- TriMed Inc.
- Location
- Santa Clarita, CA (US)
- Source
- device