DATA INNOVATIONS, LLC FDA Inspection Record
DATA INNOVATIONS, LLC is an FDA-registered facility (FEI 3001451449) located in Colchester, VT, US. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Sep 1, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 1 (1 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 6, latest Jun 14, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 1, 2016 | Voluntary Action Indicated (VAI) | Design validation - software validation not performed (21 CFR 820.30(g)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 14, 2024 | Class II | B-0514-2024 | Data Innovations Sunquest IX LIS component driver (sisixlil) v8.00.0013 used with… Data Innovations has identified an issue with Sunquest IX LIS component driver (sisixlil) v8.00.0013 used with Instrument Manager core… |
| Aug 29, 2018 | Class II | Z-0203-2019 | IM Abbott Alinity ci-series (abdicqii) accessory driver v8.00.0001 software product that… Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a… |
| Oct 23, 2014 | Class II | Z-1680-2017 | JResultNet drivers using ASTM coding language.
Calculator/data processing module, for… Software issue resulting in the results of one patient potentially being associated with another. |
| Nov 6, 2013 | Class II | Z-0571-2014 | Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100… When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003… |
| Jan 2, 2013 | Class II | B-2265-13 | Data Innovations Instrument Manager Software
Any version of Instrument Manager using the… Laboratory software, with a defect or glitch, was distributed. |
| Sep 14, 2012 | Class II | B-2299-13 | Data Innovations Instrument Manager software Model - Versions 8.12.01 Laboratory software, with a defect or glitch, was distributed. |
FDA record score
3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall B-0514-2024, this FEI as the recalling firm | Jun 14, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0203-2019, this FEI as the recalling firm | Aug 29, 2018 | 4 x 0.25 | 1 |
| Inspection classified VAI | Sep 1, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1680-2017, this FEI as the recalling firm | Oct 23, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0571-2014, this FEI as the recalling firm | Nov 6, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall B-2265-13, this FEI as the recalling firm | Jan 2, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall B-2299-13, this FEI as the recalling firm | Sep 14, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |