MADISON POLYMERIC ENGINEERING FDA Inspection Record

FEI 3001451466 · Inspections, Form 483s, warning letters and import refusals

MADISON POLYMERIC ENGINEERING is an FDA-registered facility (FEI 3001451466) located in BRANFORD, CT, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Feb 22, 2019: Voluntary Action Indicated (VAI)
Inspections on record
6 (1 OAI, 3 VAI, 2 NAI)
Last OAI inspection
Mar 23, 2015
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Apr 22, 2015. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest May 8, 2019

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Feb 22, 2019Voluntary Action Indicated (VAI)Lack of or inadequate DHR procedures (21 CFR 820.184)
Oct 3, 2017No Action Indicated (NAI)No citations listed
Apr 12, 2016Voluntary Action Indicated (VAI)Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e))
Equipment design and installation (21 CFR 820.70(g))
Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 3 more
Mar 23, 2015Official Action Indicated (OAI)Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e))
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
Changes to correct/prevent quality problems (21 CFR 820.100(a)(5))
Jan 9, 2015No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Lack of or inadequate procedures (21 CFR 820.100(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
and 1 more
Oct 26, 2011Voluntary Action Indicated (VAI)Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Quality audits - auditor independence (21 CFR 820.22)
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
Training records (21 CFR 820.25(b))

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 8, 2019Class IIZ-1924-2019Draco Enzymatic Deep-Cleaning Pad with Ready-to-Use Enzymatic Detergent, Product Number… and 1 more product
This product has been contaminated with Stenotrophomonas lactiubi.
Dec 14, 2018Class IIZ-0969-2019Draco Enzymatic Deep-Cleaning Pads, Product Number HY0305Z Product Usage: Multi-tiered…
The product is contaminated with Pseudomonas.
Mar 21, 2018Class IIZ-1895-2018First Step Endoscopic Cleaning Pad, 100ml, Flexible Endoscope Bedside Pre-Clean Kit… and 5 more products
The products exhibit an unpleasant odor due to microbial contamination with Yarrowia lipolytica.
Nov 3, 2017Class IIZ-1182-2018First Step Flexible Endoscope Bedside Pre-Clean Kit, EP-4. Intended for pre-clean of outer…
Affected lots may have exceeded its microbial limits.
Apr 17, 2017Class IIZ-2211-2017Draco Enzymatic Deep-Cleaning Pad with Enzymatic Detergent Catalogue Number: HY0305Z
Product is contaminated with Cladosporium cladosporioides/herbarum/phaenocomae.

FDA record score

7 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIFeb 22, 20196 x 0.251.5
Class II recall Z-1924-2019, this FEI as the recalling firmMay 8, 20194 x 0.251
Class II recall Z-0969-2019, this FEI as the recalling firmDec 14, 20184 x 0.251
Class II recall Z-1895-2018, this FEI as the recalling firmMar 21, 20184 x 0.251
Class II recall Z-1182-2018, this FEI as the recalling firmNov 3, 20174 x 0.251
Class II recall Z-2211-2017, this FEI as the recalling firmApr 17, 20174 x 0.251
Inspection classified VAIApr 12, 201610 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsApr 22, 201510 years or older, so it adds no points0
Inspection classified OAIMar 23, 201510 years or older, so it adds no points0
Inspection classified VAIOct 26, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing MADISON POLYMERIC ENGINEERING as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check MADISON POLYMERIC ENGINEERING yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Madison Polymeric Engineering
Location
BRANFORD, CT (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.