CONMED CORPORATION FDA Inspection Record
CONMED CORPORATION is an FDA-registered facility (FEI 3001451571) located in Utica, NY, US. RegDataLab tracks 8 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jul 13, 2022: No Action Indicated (NAI)
- Inspections on record
- 8 (7 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 22, latest Jul 14, 2026
- Import refusals
- 7
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 13, 2022 | No Action Indicated (NAI) | No citations listed |
| Sep 22, 2020 | Voluntary Action Indicated (VAI) | No citations listed |
| Feb 3, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Reporting Information Reasonably Known (21 CFR 803.50(b)(1)) |
| Feb 3, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) Process changes - review, evaluation and revalidation (21 CFR 820.75(c)) Documentation (21 CFR 820.100(b)) and 1 more |
| Jul 18, 2014 | Voluntary Action Indicated (VAI) | Equipment control activity documentation (21 CFR 820.72(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Apr 26, 2012 | Voluntary Action Indicated (VAI) | Complaints (21 CFR 820.198(a)) Documentation (21 CFR 820.100(b)) |
| Aug 1, 2011 | Voluntary Action Indicated (VAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design history file (21 CFR 820.30(j)) |
| Jan 26, 2011 | Voluntary Action Indicated (VAI) | Complaints (21 CFR 820.198(a)) Design validation - user needs and intended uses (21 CFR 820.30(g)) Documentation (21 CFR 820.100(b)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 14, 2026 | Class II | Z-3018-2026 | CONMED Linvatec GENESYS CrossFT Suture Anchor with Disposable
Driver
REF#… Suture anchor may contain incorrect patient implant card which may result in incorrect device identification and traceability. |
| May 14, 2026 | Class II | Z-2464-2026 | ConMed AirSeal 5 mm Smooth Access Port and Obturator with Blunt Tip, 150mm Length. Catalog… and 10 more products Potential for overpressure alerts. |
| Nov 18, 2022 | Class II | Z-0474-2023 | Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator… Suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the… |
| Mar 26, 2021 | Class II | Z-1498-2021 | INFINITY_TRAY Custom Procedural Kit - Canada.
containing Infinity ACL Tibial Elbow and… and 1 more product Tip Guides are potentially misaligned laterally which could affect the accuracy of the guide system and could cause a delay in procedure and… |
| Nov 12, 2020 | Class II | Z-0650-2021 | Infinity ACL Tibial Elbow Guide-indicated for use in open and arthroscopic procedures for… and 1 more product Tips of the Infinity ACL Tibial Elbow and Tip Guides are potentially misaligned laterally which could affect the accuracy of the guide… |
| Feb 23, 2018 | Class II | Z-1174-2018 | MINI INFANT < 3KG, RADIOTRANSLUCENT ELECTRODE, Universal/Anderson Connector, REF/Catalog… and 4 more products Wire sets of certain devices were assembled with adhesive on the contact wires, which may impede conductivity and could cause failure |
| Jun 19, 2017 | Class II | Z-0804-2018 | STAT 2¿ I.V. Gravity Flow Controller and STAT 2¿ Primary, Secondary Administration Sets… and 1 more product Intravascular (I.V.) fluid administration sets are exhibiting inaccurate or inconsistent flow rates. |
| Apr 17, 2017 | Class II | Z-0253-2018 | Nonabsorbable suture anchor system intended to reattach soft tissue to bone in orthopedic… Manufactured with the incorrect anchor outer body |
| Feb 28, 2017 | Class II | Z-1756-2017 | Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in… and 4 more products For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N… |
| Aug 29, 2016 | Class II | Z-0085-2017 | CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE… CONMED received complaints of damage to the CS-023 electrodes shaft insulation which could result in burn marks on the shaft. If the shaft… |
FDA record score
19 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
29 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-3018-2026, this FEI as the recalling firm | Jul 14, 2026 | 4 x 1 | 4 |
| Class II recall Z-2464-2026, this FEI as the recalling firm | May 14, 2026 | 4 x 1 | 4 |
| Class II recall Z-0474-2023, this FEI as the recalling firm | Nov 18, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Sep 22, 2020 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Feb 3, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-1498-2021, this FEI as the recalling firm | Mar 26, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0650-2021, this FEI as the recalling firm | Nov 12, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1174-2018, this FEI as the recalling firm | Feb 23, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0804-2018, this FEI as the recalling firm | Jun 19, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0253-2018, this FEI as the recalling firm | Apr 17, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1756-2017, this FEI as the recalling firm | Feb 28, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0085-2017, this FEI as the recalling firm | Aug 29, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0129-2017, this FEI as the recalling firm | Jun 27, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 3, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1842-2016, this FEI as the recalling firm | Dec 21, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0359-2016, this FEI as the recalling firm | Nov 5, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0205-2016, this FEI as the recalling firm | Jun 17, 2015 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0454-2015, this FEI as the recalling firm | Nov 6, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 18, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1719-2014, this FEI as the recalling firm | May 8, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1839-2014, this FEI as the recalling firm | Feb 20, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0331-2014, this FEI as the recalling firm | Oct 23, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1317-2013, this FEI as the recalling firm | May 3, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0826-2013, this FEI as the recalling firm | Jan 28, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0718-2013, this FEI as the recalling firm | Dec 27, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0502-2013, this FEI as the recalling firm | Nov 20, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 26, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 1, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 26, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing CONMED CORPORATION as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check CONMED CORPORATION yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- ConMed Corporation
- Location
- Utica, NY (US)
- Source
- device