Masimo Corporation FDA Inspection Record

FEI 3001452206 · Inspections, Form 483s, warning letters and import refusals

Masimo Corporation is an FDA-inspected facility (FEI 3001452206) located in Irvine, CA, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Sep 13, 2018: No Action Indicated (NAI)
Inspections on record
5 (1 OAI, 4 NAI)
Last OAI inspection
Oct 3, 2013
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Aug 13, 2014. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
3, latest Oct 28, 2015
Import refusals
1

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 13, 2018No Action Indicated (NAI)No citations listed
May 5, 2017No Action Indicated (NAI)No citations listed
Oct 3, 2013Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Lack of or inadequate procedures (21 CFR 820.100(a))
and 5 more
Feb 11, 2011No Action Indicated (NAI)No citations listed
Feb 3, 2010No Action Indicated (NAI)No citations listed

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Oct 28, 2015Class IIZ-0338-2016rainbow Reusable Sensors. RAINBOW DCI-DC3, SpO2/SpCO/SpMet, Part No. 2201; RAINBOW…
These sensors were manufactured with incompatible configurations. This could result in sensors that will provide either no readings or…
Sep 30, 2014Class IIZ-0213-2015Masimo SET uSpO2 Cable The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that…
Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed…
Mar 13, 2013Class IIZ-1187-2013Rad-8 Pulse Oximeters The Rad-8 family of pulse oximeters and accessories are…
Masimo has identified a very small number (0.02%) of Rad-8 devices with 20-pin patient cable connector that, when moved, can power off…

FDA record score

0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

5 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0338-2016, this FEI as the recalling firmOct 28, 201510 years or older, so it adds no points0
Class II recall Z-0213-2015, this FEI as the recalling firmSep 30, 201410 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsAug 13, 201410 years or older, so it adds no points0
Inspection classified OAIOct 3, 201310 years or older, so it adds no points0
Class II recall Z-1187-2013, this FEI as the recalling firmMar 13, 201310 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Masimo Corporation as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Masimo Corporation yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Masimo Corporation
Location
Irvine, CA (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.