Masimo Corporation FDA Inspection Record
Masimo Corporation is an FDA-inspected facility (FEI 3001452206) located in Irvine, CA, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Sep 13, 2018: No Action Indicated (NAI)
- Inspections on record
- 5 (1 OAI, 4 NAI)
- Last OAI inspection
- Oct 3, 2013
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Aug 13, 2014. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Oct 28, 2015
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 13, 2018 | No Action Indicated (NAI) | No citations listed |
| May 5, 2017 | No Action Indicated (NAI) | No citations listed |
| Oct 3, 2013 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Lack of or inadequate procedures (21 CFR 820.100(a)) and 5 more |
| Feb 11, 2011 | No Action Indicated (NAI) | No citations listed |
| Feb 3, 2010 | No Action Indicated (NAI) | No citations listed |
Warning letters
- Aug 13, 2014: Warning letter listed on FDA's Data Dashboard. View compliance record (420381).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 28, 2015 | Class II | Z-0338-2016 | rainbow Reusable Sensors. RAINBOW DCI-DC3, SpO2/SpCO/SpMet, Part No. 2201; RAINBOW… These sensors were manufactured with incompatible configurations. This could result in sensors that will provide either no readings or… |
| Sep 30, 2014 | Class II | Z-0213-2015 | Masimo SET uSpO2 Cable
The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that… Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed… |
| Mar 13, 2013 | Class II | Z-1187-2013 | Rad-8 Pulse Oximeters
The Rad-8 family of pulse oximeters and accessories are… Masimo has identified a very small number (0.02%) of Rad-8 devices with 20-pin patient cable connector that, when moved, can power off… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0338-2016, this FEI as the recalling firm | Oct 28, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0213-2015, this FEI as the recalling firm | Sep 30, 2014 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Aug 13, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Oct 3, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1187-2013, this FEI as the recalling firm | Mar 13, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |