Cardiac Assist, Inc FDA Inspection Record

FEI 3002228812 · Inspections, Form 483s, warning letters and import refusals

Cardiac Assist, Inc is an FDA-inspected facility (FEI 3002228812) located in Pittsburgh, PA, US. RegDataLab tracks 4 FDA inspections and 2 compliance actions at this facility.

What FDA found at this site

Latest inspection
Nov 20, 2014: No Action Indicated (NAI)
Inspections on record
4 (1 OAI, 3 NAI)
Last OAI inspection
Feb 7, 2014
Warning letters
2 listed by FDA for this FEI, latest May 13, 2014
Recalls as the recalling firm
7, latest Jun 30, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Nov 20, 2014No Action Indicated (NAI)No citations listed
Feb 7, 2014Official Action Indicated (OAI)Use and removal, Lack of or inadequate procedures (21 CFR 820.70(h))
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
Records of complaint investigation (21 CFR 820.198(e))
Quality Audits - defined intervals (21 CFR 820.22)
and 2 more
Apr 20, 2012No Action Indicated (NAI)No citations listed
Nov 13, 2009No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jun 30, 2023Class IIZ-2465-2023LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray…
Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump…
Dec 5, 2022Class IZ-0926-2023SHIP KIT, LIFESPARC, CONTROLLER AND ACCESSORIES - 5900-0000; SHIP KIT, LIFESPARC…
A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The…
Jul 21, 2022Class IZ-1763-2022TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model…
Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.
Oct 14, 2021Class IIIZ-0311-2022SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary…
Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while…
Mar 19, 2021Class IIZ-1486-2021Protek Solo Venous Dilator Set: dilator, vessel, for percutaneous catheterization…
On 02/26/2021, it was found that there were incorrectly packaged Dilator sets in Finished Goods. On 03/1/2021, it was confirmed that 14 kits…
Mar 6, 2020Class IIZ-1948-2020TandemHeart Pump Kit, Product: 5120-0000 - Product Usage: provides temporary… and 7 more products
Failure to prime due to an assembly error
Nov 19, 2019Class IIZ-1106-2020TandemLife ECLS Procedure Kit - V24/A17, Catalog Number 5740-2417 - Product Usage: The…
The kit contained a 29Fr ProtekDuo cannula instead of a 24Fr Protek Solo Venous cannula.

FDA record score

23 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 17, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

9 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-0926-2023, this FEI as the recalling firmDec 5, 202215 x 0.69
Class I recall Z-1763-2022, this FEI as the recalling firmJul 21, 202215 x 0.69
Class II recall Z-2465-2023, this FEI as the recalling firmJun 30, 20234 x 0.62.4
Class II recall Z-1486-2021, this FEI as the recalling firmMar 19, 20214 x 0.251
Class II recall Z-1948-2020, this FEI as the recalling firmMar 6, 20204 x 0.251
Class II recall Z-1106-2020, this FEI as the recalling firmNov 19, 20194 x 0.251
Warning letter on FDA's dashboard, which does not show close-outsMay 13, 201410 years or older, so it adds no points0
Inspection classified OAIFeb 7, 201410 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJun 5, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing Cardiac Assist, Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Cardiac Assist, Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Cardiac Assist, Inc
Location
Pittsburgh, PA (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.