Respironics California, LLC FDA Inspection Record
Respironics California, LLC (FEI 3002470603) located in Carlsbad, California, United States has FDA import refusal records.
What FDA found at this site
- Inspections on record
- None in FDA's published inspection data
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 14, latest Feb 28, 2022
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 28, 2022 | Class I | Z-0907-2022 | Philips Respironics V60 Ventilator (All Models, including: 1053613, 1053614, 1053615… and 2 more products Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the… |
| Jan 24, 2022 | Class I | Z-0662-2022 | Philips Respironics V60 Ventilator Part Number 1053617 and 1 more product A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially… |
| Jun 18, 2021 | Class I | Z-2073-2021 | Philips Respironics V60 Ventilator with Software Versions 3.00 and 3.10.
The… and 1 more product It has been identified that ventilators equipped with High Flow Therapy (Software Versions 3.00 and 3.10), that includes a designed safety… |
| Jan 6, 2021 | Class II | Z-1137-2021 | PHILIPS BATTERY, V60 14.4V, 11.0 Ah, 163 Wh, REF 1076374(1058272), LOT M94663-P1 - Product… Due to unqualified "design verification" batteries being shipped to customers. |
| Oct 30, 2020 | Class II | Z-0885-2021 | Philips Respironics V60 Ventilator Part Number 1053613 - Product Usage: intended to be… and 16 more products Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the… |
| Mar 5, 2020 | Class II | Z-1622-2020 | Philips V60 Ventilators with Power Management PCBA part number 1055906 Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down… |
| Sep 11, 2019 | Class II | Z-1152-2020 | Respironics V60 Ventilator, Service # 850008, REF: 1053613, 1053614, 1053617, 1053614 and 3 more products Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no… |
| Apr 24, 2017 | Class I | Z-2061-2017 | Philips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617… Over time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board… |
| Feb 1, 2017 | Class II | Z-1687-2017 | Philips V60 Ventilator with Version 2.20 Software,
Description: V
60 Ventilator,Intl… The V60 Ventilator with Version 2.20 software installed may falsely detect that the blower motor has stalled. If this condition occurs, the… |
| Oct 29, 2014 | Class II | Z-0485-2015 | Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator
The Esprit V1000 and… If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board… |
FDA record score
Not scored. FDA's published inspection data has no inspection for this site, and the score is only counted for inspected sites. This is not a score of 0. The records FDA does list for this FEI are below.
14 records FDA lists for this FEI
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0907-2022, this FEI as the recalling firm | Feb 28, 2022 | The site is not scored, so it adds no points | 0 |
| Class I recall Z-0662-2022, this FEI as the recalling firm | Jan 24, 2022 | The site is not scored, so it adds no points | 0 |
| Class I recall Z-2073-2021, this FEI as the recalling firm | Jun 18, 2021 | The site is not scored, so it adds no points | 0 |
| Class I recall Z-2061-2017, this FEI as the recalling firm | Apr 24, 2017 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-1137-2021, this FEI as the recalling firm | Jan 6, 2021 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-0885-2021, this FEI as the recalling firm | Oct 30, 2020 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-1622-2020, this FEI as the recalling firm | Mar 5, 2020 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-1152-2020, this FEI as the recalling firm | Sep 11, 2019 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-1687-2017, this FEI as the recalling firm | Feb 1, 2017 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-0485-2015, this FEI as the recalling firm | Oct 29, 2014 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0008-2015, this FEI as the recalling firm | Sep 17, 2014 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-1537-2013, this FEI as the recalling firm | Jun 4, 2013 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-2260-2012, this FEI as the recalling firm | Jul 31, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2084-2013, this FEI as the recalling firm | Oct 29, 2007 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 29, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 29, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 29, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
Reviewing Respironics California, LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Respironics California, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Respironics California, LLC → Check this supplier
Facility details
- Location
- Carlsbad, California (United States)