BRAINLAB SE FDA Inspection Record
BRAINLAB SE is an FDA-registered facility (FEI 3002619595) located in Munich Bavaria, DE. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 20, 2025: No Action Indicated (NAI)
- Inspections on record
- 4 (4 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 35, latest Feb 10, 2025
- Import refusals
- 8
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 20, 2025 | No Action Indicated (NAI) | No citations listed |
| Feb 25, 2016 | No Action Indicated (NAI) | No citations listed |
| Jun 21, 2013 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate DHR procedures (21 CFR 820.184) |
| May 27, 2011 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 10, 2025 | Class II | Z-1392-2025 | Origin Data Management software versions 3.1.0, 3.1.1, 3.1.2, 3.2.0, 3.2.1 Under certain circumstances, it can occur that patient records from unrelated individuals are automatically merged into one patient record… |
| Mar 2, 2023 | Class II | Z-1292-2023 | Cirq Arm System 2.0 Risk of mechanical instability due to potential manufacturing error of the device |
| Feb 14, 2023 | Class II | Z-1259-2023 | Brainlab ExacTrac Dynamic software, Medical Charged-Particle Radiation Therapy System… Deep Inspiration Breath Hold (DIBH) functionality may not work as specified when based on a standard-workflow Treatment Template |
| Feb 25, 2022 | Class II | Z-0733-2022 | ExacTrac Dynamic software model numbers:
20910-01, ETD Positioning and Moni. 1.0.0… The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs. |
| May 14, 2021 | Class II | Z-1846-2021 | ExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2… Display of potential patient movement might be delayed to the user for high dose treatments. |
| Mar 1, 2021 | Class II | Z-1185-2021 | ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System). In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current… |
| Aug 31, 2020 | Class II | Z-0056-2021 | Ultrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT… Brainlab Ultrasound Navigation Software does not support the modification of the probe's image width. |
| Aug 20, 2020 | Class II | Z-2925-2020 | Cirq Arm System - Model SBS 1.3.2 (Art.No. 56000-01B) and SBS 1.4 (Art.No. 56000-01C) of… Holding force of the Positioning Arm can be reduced by prolonged non-use. |
| Jul 16, 2020 | Class II | Z-2792-2020 | BRAINLAB Curve;
Model/catalogue (article) numbers for Curve: 19901; 19901B; 19900… and 1 more product Incorrect manufacturer calibration |
| Sep 25, 2019 | Class II | Z-0176-2020 | Patient Data Manager, Model/Catalogue Numbers:
25100-09 - PATIENT DATA MANAGER… The Brainlab Patient Data Manager software (PDM) in combination with the Brainlab iPlan CMF/Cranial/ENT/Spine software 2.1.1, 2.2.0, 2.2.1… |
FDA record score
25 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
35 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1392-2025, this FEI as the recalling firm | Feb 10, 2025 | 4 x 1 | 4 |
| Class I recall Z-1082-2019, this FEI as the recalling firm | Feb 27, 2019 | 15 x 0.25 | 3.8 |
| Class II recall Z-1292-2023, this FEI as the recalling firm | Mar 2, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1259-2023, this FEI as the recalling firm | Feb 14, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0733-2022, this FEI as the recalling firm | Feb 25, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1846-2021, this FEI as the recalling firm | May 14, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1185-2021, this FEI as the recalling firm | Mar 1, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0056-2021, this FEI as the recalling firm | Aug 31, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2925-2020, this FEI as the recalling firm | Aug 20, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2792-2020, this FEI as the recalling firm | Jul 16, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0176-2020, this FEI as the recalling firm | Sep 25, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1205-2019, this FEI as the recalling firm | Mar 11, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1006-2018, this FEI as the recalling firm | Feb 14, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0320-2018, this FEI as the recalling firm | Nov 16, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0172-2018, this FEI as the recalling firm | Nov 3, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0959-2017, this FEI as the recalling firm | Oct 27, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1929-2016, this FEI as the recalling firm | May 4, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0647-2016, this FEI as the recalling firm | Nov 5, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0218-2016, this FEI as the recalling firm | Sep 21, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2795-2015, this FEI as the recalling firm | Aug 18, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2440-2015, this FEI as the recalling firm | Aug 5, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0282-2016, this FEI as the recalling firm | May 8, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1582-2015, this FEI as the recalling firm | Mar 6, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1316-2015, this FEI as the recalling firm | Feb 16, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1215-2015, this FEI as the recalling firm | Jan 16, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0956-2015, this FEI as the recalling firm | Nov 19, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2134-2014, this FEI as the recalling firm | Jun 27, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2146-2014, this FEI as the recalling firm | Jun 24, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2213-2014, this FEI as the recalling firm | May 21, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1707-2014, this FEI as the recalling firm | Apr 15, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1513-2014, this FEI as the recalling firm | Feb 24, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1469-2014, this FEI as the recalling firm | Nov 18, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0967-2014, this FEI as the recalling firm | Nov 15, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2083-2013, this FEI as the recalling firm | Jul 16, 2013 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0598-2016, this FEI as the recalling firm | Apr 22, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 15, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing BRAINLAB SE as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check BRAINLAB SE yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Brainlab SE
- Location
- Munich Bavaria (DE)
- Source
- device