Siemens Healthcare Diagnostics Inc FDA Inspection Record
Siemens Healthcare Diagnostics Inc is an FDA-registered facility (FEI 3002637618) located in East Walpole, MA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 19, 2021: No Action Indicated (NAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 31, latest Apr 2, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 19, 2021 | No Action Indicated (NAI) | No citations listed |
| Aug 21, 2013 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Report of violation of the Act (see 803.52(e)(9)) (21 CFR 806.10(a)(2)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) and 1 more |
| May 19, 2010 | Voluntary Action Indicated (VAI) | Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 2, 2026 | Class II | Z-2005-2026 | epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System;
Siemens Material… Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated… |
| Jul 7, 2025 | Class II | Z-2272-2025 | epoc BGEM BUN Test Card [25pk]. Material Number: 10736515. Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are… |
| Jun 26, 2025 | Class II | Z-2230-2025 | RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097. Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error… |
| Mar 31, 2025 | Class II | Z-1673-2025 | Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma. Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in… |
| Mar 19, 2025 | Class II | Z-1684-2025 | epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material… and 1 more product The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n. |
| Sep 24, 2024 | Class II | Z-0286-2025 | epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515 Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately… |
| May 17, 2024 | Class II | Z-2214-2024 | Software version 5.3 in:
RAPIDPoint 500e Blood Gas System (USA)
Siemens Material… Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source… |
| Oct 26, 2023 | Class II | Z-0606-2024 | RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 100 -Intended for in vitro… and 3 more products RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias… |
| Oct 19, 2023 | Class II | Z-0601-2024 | epoc NXS Host Blood Analysis System, Siemens Material Numbers 11413497 (EU), 11413475… Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing… |
| Sep 26, 2023 | Class II | Z-0422-2024 | RAPIDPOINT 500e Blood Gas System and 1 more product There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results… |
FDA record score
43 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
29 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2005-2026, this FEI as the recalling firm | Apr 2, 2026 | 4 x 1 | 4 |
| Class II recall Z-2272-2025, this FEI as the recalling firm | Jul 7, 2025 | 4 x 1 | 4 |
| Class II recall Z-2230-2025, this FEI as the recalling firm | Jun 26, 2025 | 4 x 1 | 4 |
| Class II recall Z-1673-2025, this FEI as the recalling firm | Mar 31, 2025 | 4 x 1 | 4 |
| Class II recall Z-1684-2025, this FEI as the recalling firm | Mar 19, 2025 | 4 x 1 | 4 |
| Class II recall Z-0286-2025, this FEI as the recalling firm | Sep 24, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-2214-2024, this FEI as the recalling firm | May 17, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0606-2024, this FEI as the recalling firm | Oct 26, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0601-2024, this FEI as the recalling firm | Oct 19, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0422-2024, this FEI as the recalling firm | Sep 26, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-2476-2023, this FEI as the recalling firm | Jun 1, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1214-2023, this FEI as the recalling firm | Dec 12, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1372-2022, this FEI as the recalling firm | Jun 1, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-2970-2020, this FEI as the recalling firm | Aug 28, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2432-2020, this FEI as the recalling firm | May 27, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2135-2019, this FEI as the recalling firm | May 29, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2040-2017, this FEI as the recalling firm | Apr 10, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2797-2016, this FEI as the recalling firm | Aug 11, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2525-2016, this FEI as the recalling firm | Jun 22, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0609-2016, this FEI as the recalling firm | Nov 11, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2813-2015, this FEI as the recalling firm | Aug 24, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2479-2015, this FEI as the recalling firm | Jul 28, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1256-2015, this FEI as the recalling firm | Jan 22, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1252-2015, this FEI as the recalling firm | Dec 30, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2079-2014, this FEI as the recalling firm | May 23, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 21, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0635-2013, this FEI as the recalling firm | Nov 21, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0102-2014, this FEI as the recalling firm | Sep 17, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 19, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Siemens Healthcare Diagnostics Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Siemens Healthcare Diagnostics Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Siemens Healthcare Diagnostics Inc
- Location
- East Walpole, MA (US)
- Source