Arjohuntleigh Magog, Inc. FDA Inspection Record
Arjohuntleigh Magog, Inc. is an FDA-registered facility (FEI 3002655016) located in MAGOG Quebec, CA. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Oct 29, 2009: No Action Indicated (NAI)
- Inspections on record
- 1 (1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 4, latest Feb 17, 2023
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 29, 2009 | No Action Indicated (NAI) | Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 17, 2023 | Class II | Z-1629-2023 | Arjo Portable Scale Adaptor, Model Number 700.05725; accessory intended to be used in… Some Arjo Portable Scale Adaptors have been assembled with straps that have a potential to detach. |
| Mar 4, 2021 | Class II | Z-2479-2021 | Maxi Sky 1000 (Arjo) and V10 (BHM) Bariatric Ceiling Rail - Mounted Lift with measuring… The spreader bar can disconnect from the scale bottom attachment. |
| Oct 6, 2020 | Class II | Z-1100-2021 | Arjo Flat Dynamic Positioning System (DPS) - Product Usage: used in conjunction with… Risk of Sling Clip breakage while in use with obsolete Flat Dynamic Positioning System (DPS) |
| Feb 12, 2018 | Class II | Z-2660-2018 | Quick Connect 272 KG Scale used with Maxi Sky 2 ceiling lift
The Scale is intended to… and 1 more product The spreader bar can detach from the scale attachment under certain conditions. |
FDA record score
5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1629-2023, this FEI as the recalling firm | Feb 17, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-2479-2021, this FEI as the recalling firm | Mar 4, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1100-2021, this FEI as the recalling firm | Oct 6, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2660-2018, this FEI as the recalling firm | Feb 12, 2018 | 4 x 0.25 | 1 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 28, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 28, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 28, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 28, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Arjohuntleigh Magog, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Arjohuntleigh Magog, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Arjohuntleigh Magog, Inc. → Check this supplier
Facility details
- FDA Legal Name
- ARJOHUNTLEIGH MAGOG, INC.
- Company
- Arjohuntleigh Magog, Inc.
- Location
- MAGOG Quebec (CA)
- Source
- device