BioFire Diagnostics, LLC — FDA Facility Profile
BioFire Diagnostics, LLC is an FDA-registered facility (FEI 3002773840) located in Salt Lake City, UT, US. RegDataLab tracks 5 FDA inspections at this facility.
Reviewing BioFire Diagnostics, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check BioFire Diagnostics, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
What FDA found at this site
- Latest inspection
- Apr 5, 2024: No Action Indicated (NAI)
- Inspections on record
- 5 (5 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 35, latest Aug 12, 2026
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 5, 2024 | No Action Indicated (NAI) | Evaluation, timeliness, identification (21 CFR 820.198(d)) |
| Aug 5, 2021 | No Action Indicated (NAI) | No citations listed |
| Mar 18, 2016 | No Action Indicated (NAI) | No citations listed |
| Oct 18, 2012 | No Action Indicated (NAI) | No citations listed |
| Aug 5, 2010 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Periodic equipment inspections (21 CFR 820.70(g)(2)) Quality Audits - defined intervals (21 CFR 820.22) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 12, 2026 | Class II | Z-0148-2027 | Generic Name: FilmArray¿ Pneumonia Panel
Generic Name: FilmArray¿ Pneumonia… and 2 more products Due to potential contaiminated of spefic lots of pneumonia and joint infection panels that could result in false positive results |
| Jun 24, 2026 | Class II | Z-2936-2026 | BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742
UDI code: 00815381020529
PCR-based… Due to increase in false negative results |
| Apr 22, 2026 | Class II | Z-2067-2026 | BioFire Joint Infection (JI) Panel
REF: RFIT-ASY-0138, 30 test Kit. Contamination to in-vitro diagnostic test may result in false positives. |
| Apr 1, 2026 | Class II | Z-2081-2026 | Biofire Spotfire Respiratory/Sore Throat Panel
REF: 423485 Respiratory/sore throat panel test may result in false negative results and control failures. |
| Oct 22, 2025 | Class II | Z-0602-2026 | BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit) Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control… |
| Apr 30, 2025 | Class II | Z-1877-2025 | BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems Due to manufacturing error, respiratory panel may result in elevated control failures and/or false negative results. |
| Feb 27, 2025 | Class II | Z-1518-2025 | FilmArra Pneumonia Panel plus (Pneumoplus), REF: RFIT-LBL-0374 Increased risk of control failures and false negative test results with multiplexed nucleic acid test. |
| Aug 7, 2024 | Class II | Z-3122-2024 | FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W and 1 more product Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with… |
| Jun 19, 2024 | Class II | Z-2456-2024 | BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY… If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia… |
| Jun 10, 2024 | Class II | Z-2682-2024 | BIOFIRE FILMARRAY Pneumonia Panel, REF: RFIT-ASY-0144 and 1 more product Pneumonia panels have an increased risk of control failures due to an issue in the manufacturing process, which could lead to delayed… |
Facility details
- Company
- BioFire Diagnostics, LLC
- Location
- Salt Lake City, UT (US)
- Source