BIOMET UK LTD. FDA Inspection Record
BIOMET UK LTD. is an FDA-registered facility (FEI 3002806535) located in BRIDGEND, GB. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 3, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 4, latest Aug 9, 2018
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 3, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) DHR content (21 CFR 820.184) |
| Mar 18, 2015 | Voluntary Action Indicated (VAI) | Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Jul 2, 2009 | No Action Indicated (NAI) | Activity documentation (21 CFR 820.70(g)(1)) Design verification - documentation (21 CFR 820.30(f)) Design validation - documentation (21 CFR 820.30(g)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 9, 2018 | Class II | Z-3261-2018 | HipLOC Hip Fracture Fixation System- Hiploc System 130¿ Plate 38 mm Barrel 4 Hole… and 3 more products HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel |
| Jan 5, 2016 | Class II | Z-1152-2017 | Oxford Knee System Tibial Resector Body Tube & Guides
Product Usage:
Instruments for… and 23 more products Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina… |
| Mar 20, 2013 | Class II | Z-1124-2013 | REF 32-422760
Oxford Uni Knee System,Toffee Hammer, non-sterile,
found in REF… Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle… |
| Jul 27, 2012 | Class II | Z-2224-2012 | Oxford Gap Gauge SML 3/4mm.
The instrument is used for an intermediate check of the… A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model remained the same… |
FDA record score
3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Aug 3, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-3261-2018, this FEI as the recalling firm | Aug 9, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1152-2017, this FEI as the recalling firm | Jan 5, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 18, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1124-2013, this FEI as the recalling firm | Mar 20, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2224-2012, this FEI as the recalling firm | Jul 27, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |