ELEKTA LIMITED FDA Inspection Record
ELEKTA LIMITED is an FDA-registered facility (FEI 3002806900) located in CRAWLEY West Sussex, GB. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Sep 29, 2010: No Action Indicated (NAI)
- Inspections on record
- 1 (1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest May 21, 2020
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 29, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 21, 2020 | Class II | Z-2297-2020 | Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0… In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data… |
| Apr 3, 2019 | Class II | Z-1297-2019 | Elekta Unity systems
Product Usage:
Elekta Unity using Magnetic Resonance Imaging is… There is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image… |
| Mar 12, 2019 | Class II | Z-1109-2019 | Elekta Unity, Image-Guided Radiation Therapy System
Product Usage:
Elekta Unity using… Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do… |
| Jan 4, 2019 | Class II | Z-0706-2019 | Elekta Unity In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings… |
| Nov 20, 2017 | Class II | Z-0723-2018 | Agility with Integrity R3.2 and 1 more product Potential for positional errors following automatic table movement (ATM). |
FDA record score
5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2297-2020, this FEI as the recalling firm | May 21, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1297-2019, this FEI as the recalling firm | Apr 3, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1109-2019, this FEI as the recalling firm | Mar 12, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0706-2019, this FEI as the recalling firm | Jan 4, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0723-2018, this FEI as the recalling firm | Nov 20, 2017 | 4 x 0.25 | 1 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing ELEKTA LIMITED as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ELEKTA LIMITED yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Elekta Limited
- Location
- CRAWLEY West Sussex (GB)
- Source
- device