OCULUS OPTIKGERATE GMBH FDA Inspection Record
OCULUS OPTIKGERATE GMBH is an FDA-registered facility (FEI 3002807782) located in WETZLAR Hesse, DE. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Nov 10, 2016: No Action Indicated (NAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Aug 8, 2023
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 10, 2016 | No Action Indicated (NAI) | No citations listed |
| Sep 19, 2013 | Voluntary Action Indicated (VAI) | No citations listed |
| Jun 24, 2010 | Voluntary Action Indicated (VAI) | Document review, approval by designated individual (21 CFR 820.40(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 8, 2023 | Class II | Z-2482-2023 | Oculus Pentacam AXL Wave, Ref 70020, CE 0123 and 2 more products Their is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements. |
| Jul 8, 2022 | Class II | Z-1672-2022 | Pentacam AXL Wave REF 70020. Used to image the anterior segment of the eye. and 2 more products Due to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning… |
| Nov 21, 2017 | Class II | Z-0635-2018 | OCULUS Pentacam AXL, Model 70100, software versions 1.21r01, 1.21r03, 1.21r11, and… The device software versions have an anomaly which may produce an erroneous marking for the quality specification value |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2482-2023, this FEI as the recalling firm | Aug 8, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1672-2022, this FEI as the recalling firm | Jul 8, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0635-2018, this FEI as the recalling firm | Nov 21, 2017 | 4 x 0.25 | 1 |
| Inspection classified VAI | Sep 19, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 24, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 15, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing OCULUS OPTIKGERATE GMBH as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.