O-TWO MEDICAL TECHNOLOGIES, INC. FDA Inspection Record
O-TWO MEDICAL TECHNOLOGIES, INC. is an FDA-registered facility (FEI 3002807836) located in Brampton Ontario, CA. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 4, 2011: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 4, latest Feb 12, 2019
- Import refusals
- 5
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 4, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate DHR procedures (21 CFR 820.184) Investigation of device failures (21 CFR 820.198(c)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Product rework documentation, DHR {see also 820.184} (21 CFR 820.90(b)(2)) |
| Dec 11, 2008 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 12, 2019 | Class I | Z-0976-2019 | e500 Automatic Transport Ventilator, Model Number 01EVE500 and 2 more products There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper… |
| May 14, 2018 | Class II | Z-2885-2018 | O-Two Equinox¿ Relieve N2O/O2 Analgesic Gas Mixing and Delivery System
Model#: 01EQ1000 Higher levels of nitrous oxide delivered than specification |
| Sep 30, 2015 | Class II | Z-2731-2017 | Burn Relief¿, MODEL #(s): 06BK1221, 06BK1232, 06BK1201, 06BK1200, CATALOG #(s): 06BK1221… For lack of evidence to validate effectiveness of sterilization dose, O-Two Medical Technologies Inc. hereby announces a voluntary medical… |
| Jun 23, 2015 | Class II | Z-2705-2017 | O-Two Adult Ventilation Timer, MODEL #(s): 01BM1000-CS, 01BM3201-MOTP-CS… Malfunction of an O-Two Adult Ventilation Timer, where the flash rate of the malfunctioning device was reported to be twice that of the rate… |
FDA record score
5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0976-2019, this FEI as the recalling firm | Feb 12, 2019 | 15 x 0.25 | 3.8 |
| Class II recall Z-2885-2018, this FEI as the recalling firm | May 14, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2731-2017, this FEI as the recalling firm | Sep 30, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2705-2017, this FEI as the recalling firm | Jun 23, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 4, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing O-TWO MEDICAL TECHNOLOGIES, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check O-TWO MEDICAL TECHNOLOGIES, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- O-Two Medical Technologies, Inc.
- Location
- Brampton Ontario (CA)
- Source