RADIOMETER MEDICAL APS FDA Inspection Record
RADIOMETER MEDICAL APS is an FDA-registered facility (FEI 3002807968) located in BROENSHOEJ Hovedstaden, DK. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Sep 19, 2024: No Action Indicated (NAI)
- Inspections on record
- 2 (2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 18, latest Aug 18, 2026
- Import refusals
- 3
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 19, 2024 | No Action Indicated (NAI) | No citations listed |
| May 11, 2017 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 18, 2026 | Class II | Z-0190-2027 | The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained… In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types… |
| Mar 21, 2025 | Class II | Z-1659-2025 | Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090
Radiometer ABL90… ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the… |
| May 31, 2024 | Class II | Z-2426-2024 | ABL800 FLEX
Model Numbers 393-800 and 393-801
The ABL800 FLEX analyzers are intended… Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in… |
| Mar 8, 2023 | Class II | Z-1394-2023 | AQURE REF 933-599
Software Versions
2.5.2
2.5.3
2.5.4
2.6.0
2.6.1
The AQURE system… Due to potential software issue that may result in patient mix-up information. |
| May 18, 2022 | Class II | Z-1321-2022 | ABL800 Flex Analyzer Model Numbers 393-800 and 393-801 There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+. |
| Apr 8, 2022 | Class II | Z-1115-2022 | Radiometer, REF:956-552, PIC050- Arterial Blood Sampler 1 X 2 mL, 80 IU… Arterial Blood Sampler sterility cannot be guaranteed throughout the shelf life of the product. |
| Apr 14, 2021 | Class II | Z-1856-2021 | Radiometer ABL90 FLEX Analyzer with E3800 PC units.
The E3800 CPU is also used as spare… and 2 more products Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed. |
| Mar 26, 2021 | Class II | Z-1683-2021 | PIC070 Arterial Blood Sampler PICO70 - Product Usage: Preheparinized… Radiometer has become aware of an issue with PICO70 Arterial Blood Sampler. The issue relates to the sterile barrier system for the product… |
| May 26, 2020 | Class II | Z-2493-2020 | Blood gases (PCO2, PO2) and blood pH test system; ABL800 FLEX; Model Number 393-800; UDI… and 1 more product Potential risk of patient mix-up on analyzers due to software issues. |
| May 11, 2020 | Class II | Z-2467-2020 | ABL80-FLEX-393-839 Box Label, REF 393-839, IVD, CE, UDI: 05700693938394 - Product Usage… and 1 more product Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry… |
FDA record score
30 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
18 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0190-2027, this FEI as the recalling firm | Aug 18, 2026 | 4 x 1 | 4 |
| Class II recall Z-1659-2025, this FEI as the recalling firm | Mar 21, 2025 | 4 x 1 | 4 |
| Class II recall Z-2426-2024, this FEI as the recalling firm | May 31, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1394-2023, this FEI as the recalling firm | Mar 8, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1321-2022, this FEI as the recalling firm | May 18, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1115-2022, this FEI as the recalling firm | Apr 8, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1856-2021, this FEI as the recalling firm | Apr 14, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1683-2021, this FEI as the recalling firm | Mar 26, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2493-2020, this FEI as the recalling firm | May 26, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2467-2020, this FEI as the recalling firm | May 11, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2247-2020, this FEI as the recalling firm | Mar 17, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1684-2020, this FEI as the recalling firm | Mar 10, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1836-2020, this FEI as the recalling firm | Mar 6, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1755-2020, this FEI as the recalling firm | Feb 20, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1760-2020, this FEI as the recalling firm | Feb 3, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2138-2020, this FEI as the recalling firm | Nov 21, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0864-2020, this FEI as the recalling firm | Aug 8, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1582-2018, this FEI as the recalling firm | Feb 27, 2018 | 4 x 0.25 | 1 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing RADIOMETER MEDICAL APS as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check RADIOMETER MEDICAL APS yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Radiometer Medical ApS
- Location
- BROENSHOEJ Hovedstaden (DK)
- Source
- device