MAQUET CRITICAL CARE AB — FDA Facility Profile

FEI 3002808128

MAQUET CRITICAL CARE AB is an FDA-registered facility (FEI 3002808128) located in SOLNA Stockholm, SE. RegDataLab tracks 2 FDA inspections at this facility.

Reviewing MAQUET CRITICAL CARE AB as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check MAQUET CRITICAL CARE AB yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

What FDA found at this site

Latest inspection
Sep 26, 2019: Voluntary Action Indicated (VAI)
Inspections on record
2 (1 OAI, 1 VAI)
Last OAI inspection
Sep 13, 2013
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
3, latest Jun 30, 2026
Import refusals
5

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 26, 2019Voluntary Action Indicated (VAI)Lack of or inadequate DHR procedures (21 CFR 820.184)
Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2))
Lack of or inadequate process validation (21 CFR 820.75(a))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Sep 13, 2013Official Action Indicated (OAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jun 30, 2026Not yet classifiedZ-2963-2026GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N…
The module did not undergo electrical safety tests and functional tests during production.
Jun 24, 2026Not yet classifiedZ-2860-2026Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40…
Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may…
May 30, 2025Not yet classifiedZ-2013-2025Servo-u Ventilator System. Model Number: 6688600. and 2 more products
Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect…

Facility details

Company
Maquet Critical Care AB
Location
SOLNA Stockholm (SE)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.