LivaNova Deutschland GmbH FDA Inspection Record
LivaNova Deutschland GmbH is an FDA-registered facility (FEI 3002808267) located in MUNCHEN Bavaria, DE. RegDataLab tracks 2 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Apr 28, 2022: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (1 OAI, 1 VAI)
- Last OAI inspection
- Apr 13, 2011
- Warning letters
- 1 listed by FDA for this FEI, latest Aug 2, 2011
- Recalls as the recalling firm
- 7, latest Oct 18, 2024
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 28, 2022 | Voluntary Action Indicated (VAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Investigation of device failures (21 CFR 820.198(c)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Apr 13, 2011 | Official Action Indicated (OAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 18, 2024 | Class II | Z-0571-2025 | Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control… |
| Aug 3, 2023 | Class II | Z-0073-2024 | 3T Heater Cooler System, REF: 16-02-85 The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device. |
| Dec 17, 2021 | Class III | Z-0787-2022 | Sorin Centrifugal Pump System, CP5, Drive Unit (Catalogue Item #60-01-04, UDI… One ferrite ring expected to be installed on centrifugal pump system power cable may not be present. |
| Aug 26, 2019 | Class II | Z-2439-2020 | Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2… and 1 more product The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during… |
| Oct 30, 2017 | Class II | Z-0401-2018 | CP5 Centrifugal Pump, made up of the drive unit, the emergency system and the pump control… Possibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take… |
| Dec 4, 2013 | Class II | Z-0856-2014 | S5/C5 Heart-lung machine
Product Usage: The StOckert S5/ Sorin C5 System is intended… and 1 more product Sorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob… |
| Sep 24, 2013 | Class II | Z-0142-2014 | S5 Heart-lung machine, Stockert S5 System. Catalogue No: 48-30/40/50-00.
The Stockert… Sorin Group is recalling certain models of S5 Perfusion Systems due to malfunction of cardioplegia control. |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Apr 28, 2022 | 6 x 0.6 | 3.6 |
| Class II recall Z-0571-2025, this FEI as the recalling firm | Oct 18, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0073-2024, this FEI as the recalling firm | Aug 3, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-2439-2020, this FEI as the recalling firm | Aug 26, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0401-2018, this FEI as the recalling firm | Oct 30, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0856-2014, this FEI as the recalling firm | Dec 4, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0142-2014, this FEI as the recalling firm | Sep 24, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Aug 2, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Apr 13, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 14, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 2, 2026 | Newest record Oct 22, 2026 | Current |
Reviewing LivaNova Deutschland GmbH as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.