Electro-Diagnostic Imaging, Inc. FDA Inspection Record
Electro-Diagnostic Imaging, Inc. is an FDA-inspected facility (FEI 3002900377) located in Milpitas, CA, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- May 26, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (3 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 26, 2017 | Voluntary Action Indicated (VAI) | Design validation - software validation documentation (21 CFR 820.30(g)) Design input - documentation (21 CFR 820.30(c)) Design history file (21 CFR 820.30(j)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 2 more |
| Jun 26, 2013 | No Action Indicated (NAI) | No citations listed |
| Nov 3, 2011 | Voluntary Action Indicated (VAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design validation - software validation not performed (21 CFR 820.30(g)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
| Feb 22, 2010 | Voluntary Action Indicated (VAI) | Documentation of software validation (21 CFR 820.70(i)) Retention period (21 CFR 820.180(b)) Uniform and timely processing (21 CFR 820.198(a)(1)) Processing MDRs {see also 803, 804} (21 CFR 820.198(a)(3)) and 3 more |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
3 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | May 26, 2017 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Nov 3, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 22, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Electro-Diagnostic Imaging, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Electro-Diagnostic Imaging, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Electro-Diagnostic Imaging, Inc.
- Location
- Milpitas, CA (US)
- Source
- inspections