Neurotherm LLC FDA Inspection Record
Neurotherm LLC is an FDA-inspected facility (FEI 3002953813) located in Wilmington, MA, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Feb 4, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (3 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jul 11, 2012. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Oct 9, 2013
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 4, 2016 | Voluntary Action Indicated (VAI) | Report of risk to health (21 CFR 806.10(a)(1)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Jan 9, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) and 9 more |
| Apr 26, 2012 | No Action Indicated (NAI) | No citations listed |
| Dec 4, 2008 | Voluntary Action Indicated (VAI) | No citations listed |
Warning letters
- Jul 11, 2012: Warning letter listed on FDA's Data Dashboard. View compliance record (306318).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 9, 2013 | Class II | Z-1361-2016 | Neurotherm NT2000iX RF Generator
Model Number: RFG-NT -2000iX NT2000iX software shipped with the international setting turned on. The international settings include access to Corodotomy, Bi-Polar, and… |
| Feb 26, 2013 | Class II | Z-1247-2016 | Neurotherm Curved Radiofrequency (RF) cannnula 10 cm
18 gauge, 10 mm active tip
Model… Straight needle labeled as a curved needle |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Feb 4, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1361-2016, this FEI as the recalling firm | Oct 9, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1247-2016, this FEI as the recalling firm | Feb 26, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 9, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jul 11, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 4, 2008 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |