DEROYAL INDUSTRIES, INC. FDA Inspection Record
DEROYAL INDUSTRIES, INC. is an FDA-registered facility (FEI 3003040625) located in NEW TAZEWELL, TN, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jul 25, 2017: No Action Indicated (NAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Sep 17, 2013
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 25, 2017 | No Action Indicated (NAI) | No citations listed |
| Aug 26, 2010 | Voluntary Action Indicated (VAI) | Procedures not adequately established or maintained (21 CFR 820.40) Buildings (21 CFR 820.70(f)) Complaints (21 CFR 820.198(a)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) and 2 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 17, 2013 | Class II | Z-0834-2014 | Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries… and 5 more products The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
2 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0834-2014, this FEI as the recalling firm | Sep 17, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 26, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing DEROYAL INDUSTRIES, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check DEROYAL INDUSTRIES, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- DeRoyal Industries Inc
- Location
- NEW TAZEWELL, TN (US)
- Source