Biosense Webster, Inc. FDA Inspection Record
Biosense Webster, Inc. is an FDA-inspected facility (FEI 3003113629) located in Irwindale, CA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Sep 27, 2016: No Action Indicated (NAI)
- Inspections on record
- 3 (3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 17, latest Jun 29, 2022
- Import refusals
- 3
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 27, 2016 | No Action Indicated (NAI) | No citations listed |
| Sep 25, 2015 | No Action Indicated (NAI) | No citations listed |
| Oct 5, 2012 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 29, 2022 | Class II | Z-1677-2022 | CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath, REF: D138501. Used to introduce various… Outside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have… |
| May 7, 2018 | Class I | Z-2878-2018 | WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523… When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force… |
| Sep 6, 2017 | Class II | Z-0054-2018 | THERMOCOOL SMARTTOUCH SF Uni-Directional Product Codes D-1347-01-S, D-1347-02-S… and 4 more products Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for… |
| Jun 14, 2017 | Class II | Z-0210-2018 | SMARTABLATE RF System Generator Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets. |
| May 17, 2017 | Class II | Z-0211-2018 | PENTARAY NAV High-Density Mapping Catheter. and 1 more product Contraindication language updated in the Instructions For Use (IFU) to indicate that Pentaray Catheters should not be used in patients with… |
| Aug 15, 2016 | Class II | Z-2776-2016 | CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C… Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China… |
| Apr 14, 2016 | Class II | Z-1590-2016 | Pentaray NAV High-Density Mapping Catheter, Catalog No. D128201, D128202, D128203… and 1 more product Biosense Webster is recalling the Pentaray Catheters to clarify the contraindication language in the Instructions For Use (IFU) and product… |
| Sep 10, 2015 | Class II | Z-0002-2016 | Carto 3 EP Navigation System. Electro physiology system which views of the electrical… and 1 more product Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was… |
| Apr 17, 2015 | Class II | Z-1525-2015 | WEBSTER HIS Catheter, 4 Pole Fixed Curve with Auto ID Technology, for mapping of cardiac… The Webster HIS catheter, 4Pole Fixed Curve with Auto ID, is not being recognized by the CARTO 3 navigation System. |
| Mar 19, 2015 | Class II | Z-1691-2015 | SMARTABLATE RF Generator System, Catalog No. M490006, M490007, M4900106, and… Biosense Webster is recalling the SMARTABLATE Generator System because Biosense Webster has received two complaints reporting that during an… |
FDA record score
9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
17 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-2878-2018, this FEI as the recalling firm | May 7, 2018 | 15 x 0.25 | 3.8 |
| Class II recall Z-1677-2022, this FEI as the recalling firm | Jun 29, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0054-2018, this FEI as the recalling firm | Sep 6, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0210-2018, this FEI as the recalling firm | Jun 14, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0211-2018, this FEI as the recalling firm | May 17, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2776-2016, this FEI as the recalling firm | Aug 15, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1590-2016, this FEI as the recalling firm | Apr 14, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0002-2016, this FEI as the recalling firm | Sep 10, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1525-2015, this FEI as the recalling firm | Apr 17, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1691-2015, this FEI as the recalling firm | Mar 19, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1242-2015, this FEI as the recalling firm | Feb 25, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0139-2015, this FEI as the recalling firm | Oct 19, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1176-2014, this FEI as the recalling firm | Feb 20, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1045-2014, this FEI as the recalling firm | Feb 12, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1617-2013, this FEI as the recalling firm | Jun 19, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1023-2013, this FEI as the recalling firm | Mar 5, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2369-2012, this FEI as the recalling firm | Nov 17, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Biosense Webster, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Biosense Webster, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Biosense Webster, Inc.
- Location
- Irwindale, CA (US)
- Source
- inspections