ZOE MEDICAL, INC. FDA Inspection Record
ZOE MEDICAL, INC. is an FDA-registered facility (FEI 3003294644) located in TOPSFIELD, MA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- May 31, 2024: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (3 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Aug 13, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 31, 2024 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Design validation - production units (21 CFR 820.30(g)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) and 1 more |
| Sep 20, 2019 | No Action Indicated (NAI) | Documented evaluation (21 CFR 820.50(a)(1)) |
| Oct 12, 2016 | No Action Indicated (NAI) | No citations listed |
| Feb 7, 2013 | Voluntary Action Indicated (VAI) | Equipment Installation, Placement, Specified Requirements (21 CFR 820.70(g)) |
| Apr 20, 2010 | Voluntary Action Indicated (VAI) | DMR required activities {see also 820.181} (21 CFR 820.80(d)(1)) Documentation (21 CFR 820.100(b)) Records not kept (21 CFR 806.20(a)) Production and Process Change Procedures, lack of or Inad. (21 CFR 820.70(b)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 13, 2024 | Class II | Z-3098-2024 | Spacelabs Ultraview DM3 Vital Signs Monitor (all variations -M, -MT, -N, -NT), Model… and 1 more product This product labeling does not adequately provide warning about the risks associated with the use of damaged batteries or batteries past the… |
| Jun 6, 2012 | Class II | Z-2380-2012 | Spacelabs Ultraview DM3 Monitor
Model Number: 91330-NT
The Ultraview DM3 Monitor is… Unit fails to power up, resulting in an equipment alarm |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | May 31, 2024 | 6 x 0.6 | 3.6 |
| Class II recall Z-3098-2024, this FEI as the recalling firm | Aug 13, 2024 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Feb 7, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2380-2012, this FEI as the recalling firm | Jun 6, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 20, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing ZOE MEDICAL, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ZOE MEDICAL, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Zoe Medical Incorporated
- Location
- TOPSFIELD, MA (US)
- Source
- device