Collagen Matrix, Inc FDA Inspection Record

FEI 3003300673 · Inspections, Form 483s, warning letters and import refusals

Collagen Matrix, Inc is an FDA-registered facility (FEI 3003300673) located in Allendale, NJ, US. RegDataLab tracks 4 FDA inspections and 2 compliance actions at this facility.

What FDA found at this site

Latest inspection
Jul 13, 2017: No Action Indicated (NAI)
Inspections on record
4 (1 OAI, 2 VAI, 1 NAI)
Last OAI inspection
Nov 25, 2015
Warning letters
2 listed by FDA for this FEI, latest Sep 20, 2016
Recalls as the recalling firm
6, latest Dec 13, 2021

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jul 13, 2017No Action Indicated (NAI)No citations listed
Nov 25, 2015Official Action Indicated (OAI)Design validation - user needs and intended uses (21 CFR 820.30(g))
Report of violation of the Act (see 803.52(e)(9)) (21 CFR 806.10(a)(2))
Dec 5, 2011Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Complaints (21 CFR 820.198(a))
Documented evaluation (21 CFR 820.50(a)(1))
Nov 10, 2009Voluntary Action Indicated (VAI)Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Design validation - production units (21 CFR 820.30(g))
Nature and details of complaint (21 CFR 820.198(e)(5))
CAPA verification/validation of corrective/preventive action (21 CFR 820.100(a)(4))
and 1 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Dec 13, 2021Class IIZ-0668-2022DuraMatrix-Onlay Collagen Dura Substitute Membrane 4 inches x 5 inches (10.0cm x… and 1 more product
Outer packaging (outer pouch) is not sealed compromising sterility and result in potential risk of patient infection, which could lead to…
Mar 19, 2019Class IIZ-1389-2019Neomem Resorbable Collagen Membrane, 15 mm x 20 mm Reference Number (model): N1520…
Product was packaged in the wrong box; may have been placed inside a Neomem FLEXPLUS box instead of a Neomem box
Aug 16, 2018Class IIIZ-0414-2019BioMend Extend, REF# 0142Z, Synthetic bone grafting material.
There is a possibility that the product was packaged in the wrong box. Specifically, the product may have been placed inside a BioMend box…
Nov 17, 2015Class IIIZ-1173-2016ZCORE Porcine Xenograft Particulate in Syringe Product Intended for use in dental…
The distributor, Osteogenics Biomedical, notified the firm that the syringes were not working properly and the syringes were being held in…
Aug 25, 2015Class IIZ-1096-2016NuOSS Cancellous Intended for use in dental surgery.
On 8/25/2015, Collagen Matrix, Inc discovered that one of the 171 units of NuOss Cancellous product shipped to a distributor may have…
May 13, 2013Class IIZ-1097-2016NuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm…
During the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf life designated for the…

FDA record score

3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 17, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

9 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0668-2022, this FEI as the recalling firmDec 13, 20214 x 0.62.4
Class II recall Z-1389-2019, this FEI as the recalling firmMar 19, 20194 x 0.251
Warning letter on FDA's dashboard, which does not show close-outsSep 20, 201610 years or older, so it adds no points0
Inspection classified OAINov 25, 201510 years or older, so it adds no points0
Class II recall Z-1096-2016, this FEI as the recalling firmAug 25, 201510 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJan 8, 201510 years or older, so it adds no points0
Class II recall Z-1097-2016, this FEI as the recalling firmMay 13, 201310 years or older, so it adds no points0
Inspection classified VAIDec 5, 201110 years or older, so it adds no points0
Inspection classified VAINov 10, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing Collagen Matrix, Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Collagen Matrix, Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Collagen Matrix, Inc.
Location
Allendale, NJ (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.