MEDIVANCE, INC. FDA Inspection Record
MEDIVANCE, INC. is an FDA-registered facility (FEI 3003329224) located in LOUISVILLE, CO, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jul 20, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (1 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Jun 27, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 20, 2016 | Voluntary Action Indicated (VAI) | Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) |
| Nov 1, 2012 | No Action Indicated (NAI) | No citations listed |
| Jan 16, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 27, 2024 | Class II | Z-2457-2024 | Arctic Sun Temperature Management Systems, Name/REF:
Arctic Sun STAT… If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to… |
| Feb 16, 2024 | Class II | Z-1381-2024 | Medivance Neonatal ArcticGel Pads, REF: 31802, 3180202, 31802C Neonatal pads are experiencing reduced water flow, which may reduce or prevent water heating or cooling performance of the pads that are… |
FDA record score
5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
3 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2457-2024, this FEI as the recalling firm | Jun 27, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1381-2024, this FEI as the recalling firm | Feb 16, 2024 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Jul 20, 2016 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing MEDIVANCE, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check MEDIVANCE, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Medivance Inc.
- Location
- LOUISVILLE, CO (US)
- Source
- device