Fresenius Medical Care Renal Therapies Group, LLC FDA Inspection Record
Fresenius Medical Care Renal Therapies Group, LLC is an FDA-registered facility (FEI 3003464075) located in LAWRENCE, MA, US. Registered operations: ANALYSIS. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 10, 2016: No Action Indicated (NAI)
- Inspections on record
- 3 (3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 16, latest Oct 14, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 10, 2016 | No Action Indicated (NAI) | No citations listed |
| Jan 8, 2013 | No Action Indicated (NAI) | No citations listed |
| May 6, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 14, 2026 | Class II | D-0167-2027 | DELFLEX Peritoneal Dialysis Solution with 1.5% Dextrose, Low Magnesium/ Low Calcium and… and 4 more products Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the… |
| Apr 28, 2025 | Class II | Z-1864-2025 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404… Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face… |
| May 14, 2024 | Class II | Z-2160-2024 | NxStage PureFlow-B Solution, Premixed Dialysate for Hemodialysis, Device Catalog Numbers… The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte… |
| Jan 29, 2024 | Class II | Z-1191-2024 | NxStage Cartridge Express with Speedswap, Catalog No. CAR-535.
An accessory device for… Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient… |
| Aug 27, 2019 | Class II | Z-0697-2020 | Low Volume Cartridge IFU (SKU CAR-125-B) used with the following devices:
1) NxStage… There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low… |
| Apr 11, 2019 | Class II | Z-1706-2019 | NxStage PureFlow B Solution, REF RFP-400, M535RFP4000
Product NxStage PureFlow B… and 1 more product Certain lots were mislabeled with incorrect product number. |
| Oct 17, 2018 | Class II | Z-0500-2019 | NxStage PureFlow B Solution-RFP-400, Premixed Dialysate for Hemodialysis.
NxStage… and 7 more products PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the… |
| Jun 11, 2018 | Class II | Z-2793-2018 | NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A)
Product… There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time… |
| Jan 29, 2018 | Class II | Z-0811-2018 | NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3)… May contain endotoxin levels which have been confirmed to exceed the ANSI/AAMI quality standard dialysis fluid of 0.5 EU/ml when used to… |
| Nov 10, 2017 | Class II | Z-1132-2018 | NxStage PureFlow B Solution - Premixed Dialysate for Hemodialysis with NxStage System One Using different fluid formulations of NxStage PureFlow B Solution at the same time during treatment may cause patient health risk. The… |
FDA record score
19 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
16 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall D-0167-2027, this FEI as the recalling firm | Oct 14, 2026 | 4 x 1 | 4 |
| Class II recall Z-1864-2025, this FEI as the recalling firm | Apr 28, 2025 | 4 x 1 | 4 |
| Class II recall Z-2160-2024, this FEI as the recalling firm | May 14, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1191-2024, this FEI as the recalling firm | Jan 29, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0697-2020, this FEI as the recalling firm | Aug 27, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1706-2019, this FEI as the recalling firm | Apr 11, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0500-2019, this FEI as the recalling firm | Oct 17, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2793-2018, this FEI as the recalling firm | Jun 11, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0811-2018, this FEI as the recalling firm | Jan 29, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1132-2018, this FEI as the recalling firm | Nov 10, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0326-2016, this FEI as the recalling firm | Oct 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0335-2016, this FEI as the recalling firm | Oct 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1854-2014, this FEI as the recalling firm | May 14, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1073-2014, this FEI as the recalling firm | Jan 24, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1214-2013, this FEI as the recalling firm | Mar 18, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0714-2013, this FEI as the recalling firm | Dec 20, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Fresenius Medical Care Renal Therapies Group, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Fresenius Medical Care Renal Therapies Group, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Fresenius Medical Care Renal Therapies Group, LLC
- Location
- LAWRENCE, MA (US)
- Registration Expires
- Dec 31, 2027
- Operations
- ANALYSIS
- Source
- both