Interventional Spine Inc FDA Inspection Record
Interventional Spine Inc is an FDA-inspected facility (FEI 3003523239) located in Irvine, CA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Sep 6, 2016: No Action Indicated (NAI)
- Inspections on record
- 3 (3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Nov 22, 2016
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 6, 2016 | No Action Indicated (NAI) | No citations listed |
| Oct 14, 2011 | No Action Indicated (NAI) | No citations listed |
| Feb 4, 2011 | No Action Indicated (NAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1)) Personnel (21 CFR 820.25(a)) Report of risk to health (21 CFR 806.10(a)(1)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 22, 2016 | Class II | Z-0941-2017 | G3 Opticage Expandable Interbody Fusion Device, Model No. 9180-00, 9180-01, 9180-06… During implantation, the proximal end of the device could separate if the angle between the device and insertion instrument is too steep… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
1 record behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0941-2017, this FEI as the recalling firm | Nov 22, 2016 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Interventional Spine Inc as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Interventional Spine Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Interventional Spine Inc → Check this supplier
Facility details
- Company
- Interventional Spine Inc
- Location
- Irvine, CA (US)
- Source
- inspections