QIAGEN FDA Inspection Record
QIAGEN is an FDA-registered facility (FEI 3003572099) located in Germantown, MD, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jul 16, 2021: Voluntary Action Indicated (VAI)
- Inspections on record
- 8 (1 OAI, 3 VAI, 4 NAI)
- Last OAI inspection
- Feb 5, 2016
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest May 16, 2016. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 18, latest Mar 12, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 16, 2021 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Feb 21, 2018 | Voluntary Action Indicated (VAI) | No citations listed |
| Dec 19, 2016 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Sampling methods - Lack of or inadequate procedures (21 CFR 820.250(b)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Feb 5, 2016 | Official Action Indicated (OAI) | Design output - Lack of or inadequate procedures (21 CFR 820.30(d)) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) and 11 more |
| Mar 26, 2015 | No Action Indicated (NAI) | No citations listed |
| Jun 26, 2014 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Feb 18, 2014 | No Action Indicated (NAI) | No citations listed |
| Sep 8, 2010 | No Action Indicated (NAI) | No citations listed |
Warning letters
- May 16, 2016: Warning letter listed on FDA's Data Dashboard. View compliance record (490174).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Mar 12, 2025 | Class II | Z-1504-2025 | Brand Name: QIAstat-Dx
Product Name: QIAstat-Dx Respiratory Panel Plus
Reference Number… Identified faulty cartridges. If such a faulty cartridge is used for sample testing, false test results could occur. |
| Jun 3, 2024 | Class II | Z-2353-2024 | QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and… Heater shaker module does not perform heating if the temperature is set below 40¿C for static incubation steps when running IVD protocol… |
| Mar 4, 2024 | Class II | Z-1508-2024 | EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic… Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE… |
| Apr 28, 2023 | Class II | Z-1683-2023 | QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test… Faulty cartridges in the lot could result in false test results. |
| Jul 20, 2022 | Class II | Z-1650-2022 | therascreen FGFR RGQ RT-PCR kit (US IVD)
REF 874721 and 2 more products Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported… |
| Feb 25, 2022 | Class II | Z-0951-2022 | therascreen KRAS RGQ PCR Kit (24)
Model Number(s): REF 870021 (US IVD) and 1 more product False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the… |
| Nov 21, 2021 | Class II | Z-1746-2022 | QIAcube Connect MDx, Model No. 9003070 During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated… |
| Nov 8, 2021 | Class II | Z-0455-2022 | QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223 The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an… |
| Jan 15, 2021 | Class II | Z-0941-2021 | SARS-CoV-2 Antigen Test QIAGEN has become aware of the potential for false positive results to occur with some patient samples. |
| Dec 11, 2020 | Class II | Z-0937-2021 | therascreen PIK3CA RGQ PCR Kit (23) US IVD product- aid clinicians in identifying breast… Kit may generate false Q546R mutation positive results caused by nonspecific molecular interactions within the Q546R higher than previously… |
FDA record score
33 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
23 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1504-2025, this FEI as the recalling firm | Mar 12, 2025 | 4 x 1 | 4 |
| Class I recall Z-2088-2019, this FEI as the recalling firm | Jun 5, 2019 | 15 x 0.25 | 3.8 |
| Class II recall Z-2353-2024, this FEI as the recalling firm | Jun 3, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1508-2024, this FEI as the recalling firm | Mar 4, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1683-2023, this FEI as the recalling firm | Apr 28, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1650-2022, this FEI as the recalling firm | Jul 20, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0951-2022, this FEI as the recalling firm | Feb 25, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Jul 16, 2021 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Feb 21, 2018 | 6 x 0.25 | 1.5 |
| Class II recall Z-1746-2022, this FEI as the recalling firm | Nov 21, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0455-2022, this FEI as the recalling firm | Nov 8, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0941-2021, this FEI as the recalling firm | Jan 15, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0937-2021, this FEI as the recalling firm | Dec 11, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2938-2020, this FEI as the recalling firm | Jul 15, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2290-2020, this FEI as the recalling firm | Apr 8, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1636-2020, this FEI as the recalling firm | Feb 21, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1476-2020, this FEI as the recalling firm | Jan 28, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2319-2019, this FEI as the recalling firm | Jul 2, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0383-2019, this FEI as the recalling firm | Jul 23, 2018 | 4 x 0.25 | 1 |
| Warning letter on FDA's dashboard, which does not show close-outs | May 16, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Feb 5, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1230-2016, this FEI as the recalling firm | Feb 2, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 26, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing QIAGEN as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check QIAGEN yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Qiagen Sciences LLC
- Location
- Germantown, MD (US)
- Source
- device