LIN-ZHI INTERNATIONAL, INC. FDA Inspection Record

FEI 3003610499 · Inspections, Form 483s, warning letters and import refusals

LIN-ZHI INTERNATIONAL, INC. is an FDA-registered facility (FEI 3003610499) located in Santa Clara, CA, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Sep 3, 2015: Voluntary Action Indicated (VAI)
Inspections on record
3 (2 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
4, latest Jan 11, 2019

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 3, 2015Voluntary Action Indicated (VAI)Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Lack of or inadequate procedures (21 CFR 820.100(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1))
May 15, 2012Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2))
Nov 22, 2010No Action Indicated (NAI)Documentation (21 CFR 820.100(b))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jan 11, 2019Class IIIZ-1256-2019Ethyl Alcohol Enzymatic Assay, (a) REF 0220 (small test kit), (b) REF 0221 (large test…
Shelf life of the product may be reduced due to degradation of the assay.
Feb 18, 2014Class IIIZ-1270-2014Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in…
Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff…
Feb 6, 2014Class IIZ-1178-2014Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from… and 1 more product
Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give…
Oct 3, 2013Class IIIZ-0386-2014Single Analyte Urine DAU Control (5 m L Norbuprenorphine, Level II control For In Vitro…
Product catalog # A68825 Norbuprenorphine 13 ng/mL Level 2 Controls gives a decreased concentration reading due to the degradation of the…

FDA record score

0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 24, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

3 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAISep 3, 201510 years or older, so it adds no points0
Class II recall Z-1178-2014, this FEI as the recalling firmFeb 6, 201410 years or older, so it adds no points0
Inspection classified VAIMay 15, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 24, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 24, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing LIN-ZHI INTERNATIONAL, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check LIN-ZHI INTERNATIONAL, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Lin-Zhi International Inc
Location
Santa Clara, CA (US)
Source
device