Theralase Technologies Inc. FDA Inspection Record
Theralase Technologies Inc. is an FDA-registered facility (FEI 3003614490) located in Toronto Ontario, CA. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Nov 15, 2012: Official Action Indicated (OAI)
- Inspections on record
- 1 (1 OAI)
- Last OAI inspection
- Nov 15, 2012
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 6, latest Sep 26, 2019
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 15, 2012 | Official Action Indicated (OAI) | Design control - no procedures (21 CFR 820.30(a)) Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Supplier notification of changes (21 CFR 820.50(b)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 26, 2019 | Class III | Z-0606-2020 | TLC-2000 Therapeutic Medical Laser System, used for the temporary relief of knee pain. and 1 more product Device was shipped without a Unique Device Identification (UDI) label. |
| Sep 26, 2019 | Class III | Z-1266-2020 | TLC-2000 Therapeutic Medical Laser System Laser Probe (Model Number: TLC-2001) Electromagnetic Interference ( EMI ) in excess of IEC-60601-1-2 (4th edition) standard |
| Sep 26, 2019 | Class III | Z-1326-2020 | TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary… Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2… |
| Sep 26, 2019 | Class III | Z-1348-2020 | TLC-2000 Therapeutic Medical Laser System Power Pack, Model Number TLC-2002 Power Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however… |
| Sep 26, 2019 | Class III | Z-1353-2020 | TLC-2000 Therapeutic Medical Laser System 660 nm laser diodes may be below rated power output. |
| Sep 26, 2019 | Class III | Z-1354-2020 | TLC-2000 Therapeutic Medical Laser System 905 nm laser diodes may be below rated power output. |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
1 record behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified OAI | Nov 15, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Theralase Technologies Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Theralase Technologies Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Theralase Inc.
- Location
- Toronto Ontario (CA)
- Source
- device