Syncardia Systems, LLC FDA Inspection Record
Syncardia Systems, LLC is an FDA-registered facility (FEI 3003761017) located in TUCSON, AZ, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Nov 29, 2018: No Action Indicated (NAI)
- Inspections on record
- 5 (3 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Apr 3, 2018. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Aug 1, 2023
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 29, 2018 | No Action Indicated (NAI) | No citations listed |
| Aug 18, 2017 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
| Apr 17, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) |
| Mar 28, 2013 | No Action Indicated (NAI) | No citations listed |
| Sep 23, 2011 | Voluntary Action Indicated (VAI) | Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) |
Warning letters
- Apr 3, 2018: Warning letter listed on FDA's Data Dashboard. View compliance record (537226).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 1, 2023 | Class II | Z-0126-2024 | SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy. |
| Nov 2, 2022 | Class II | Z-1022-2023 | SynCardia Total Artificial Heart System (70CC): 500101-001
SynCardia Total Artificial… Due to tears reported tears in the cannula due to wear and tear stress. |
| Oct 21, 2016 | Class I | Z-0659-2017 | 5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart… The Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning… |
| Aug 6, 2015 | Class I | Z-2731-2015 | Temporary Total Artificial Heart (TAH-t) with Freedom Driver System
REF… SynCardia Systems, Inc. is recalling Freedom Drivers due to a specific component of the drive mechanism may fail and cause the drive… |
| Dec 5, 2012 | Class II | Z-0620-2013 | Companion External Battery, Part Number: 293001-001
The Companion External Battery is a… Syncardia Systems is recalling the Companion External Batteries used with the Companion 2 Driver System because they may potentially be at… |
FDA record score
14 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Warning letter on FDA's dashboard, which does not show close-outs | Apr 3, 2018 | 30 x 0.25 | 7.5 |
| Class II recall Z-0126-2024, this FEI as the recalling firm | Aug 1, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1022-2023, this FEI as the recalling firm | Nov 2, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Aug 18, 2017 | 6 x 0.25 | 1.5 |
| Class I recall Z-0659-2017, this FEI as the recalling firm | Oct 21, 2016 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-2731-2015, this FEI as the recalling firm | Aug 6, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 17, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0620-2013, this FEI as the recalling firm | Dec 5, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 23, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Syncardia Systems, LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Syncardia Systems, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Syncardia Systems, LLC → Check this supplier
Facility details
- FDA Legal Name
- SYNCARDIA SYSTEMS, LLC
- Company
- SynCardia Systems LLC
- Location
- TUCSON, AZ (US)
- Source
- device