Philips Medical Systems DMC GmbH FDA Inspection Record
Philips Medical Systems DMC GmbH is an FDA-registered facility (FEI 3003768251) located in HAMBURG, DE. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 19, 2023: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (1 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 7, latest Oct 1, 2025
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 19, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Report of violation of the Act (see 803.52(e)(9)) (21 CFR 806.10(a)(2)) |
| Jul 9, 2015 | No Action Indicated (NAI) | No citations listed |
| Jun 14, 2013 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 1, 2025 | Class II | Z-0461-2026 | ProxiDiagnost N90;
Model Number: 706110; and 5 more products Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to… |
| May 19, 2021 | Class II | Z-1325-2022 | ProxiDiagnost N90. radiography and fluoroscopy system and 4 more products Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an… |
| Jul 22, 2020 | Class II | Z-2357-2021 | CombiDiagnost R90
Software Version R1.0 and R1.1 Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter… |
| Dec 19, 2019 | Class II | Z-1328-2020 | CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system… and 1 more product When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the… |
| Nov 7, 2018 | Class II | Z-0612-2019 | Stationary Fluoroscopic X-ray System CombiDiagnost R90, Model 709030
Stationary… kV/mA lockin not functioning as specified which may result in the patient receiving an increased radiation dose. |
| Sep 19, 2018 | Class II | Z-0239-2019 | Philips DigitalDiagnost software 4.1.x and 4.2.x, A Stationary X-ray System
Models… Stitching run on SkyPlate detector interrupted intermittently; Stitching with SkyPlate aborts after first image; System software CD… |
| Jun 27, 2018 | Class II | Z-2841-2018 | DigitalDiagnost 2.x Upgrades (Stitching Patient Support) 712082 and 27 more products The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator… |
FDA record score
14 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0461-2026, this FEI as the recalling firm | Oct 1, 2025 | 4 x 1 | 4 |
| Inspection classified VAI | Jan 19, 2023 | 6 x 0.6 | 3.6 |
| Class II recall Z-1325-2022, this FEI as the recalling firm | May 19, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2357-2021, this FEI as the recalling firm | Jul 22, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1328-2020, this FEI as the recalling firm | Dec 19, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0612-2019, this FEI as the recalling firm | Nov 7, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0239-2019, this FEI as the recalling firm | Sep 19, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2841-2018, this FEI as the recalling firm | Jun 27, 2018 | 4 x 0.25 | 1 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 15, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing Philips Medical Systems DMC GmbH as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.