GULF FIBEROPTICS, INC. FDA Inspection Record

FEI 3003768921 · Inspections, Form 483s, warning letters and import refusals

GULF FIBEROPTICS, INC. is an FDA-registered facility (FEI 3003768921) located in OLDSMAR, FL, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Mar 23, 2023: No Action Indicated (NAI)
Inspections on record
8 (5 VAI, 3 NAI)
Last OAI inspection
None on record
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jul 28, 2011. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
1, latest Apr 9, 2012

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 23, 2023No Action Indicated (NAI)No citations listed
Dec 18, 2019Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
Lack of or inadequate process validation (21 CFR 820.75(a))
Aug 13, 2015No Action Indicated (NAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
Documentation (21 CFR 820.80(c))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
and 1 more
May 6, 2013Voluntary Action Indicated (VAI)Documentation (21 CFR 820.80(e))
Nonconforming product control (21 CFR 820.90(a))
Lack of or inadequate DHR procedures (21 CFR 820.184)
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Feb 3, 2012Voluntary Action Indicated (VAI)Design input - documentation (21 CFR 820.30(c))
Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a))
Documentation (21 CFR 820.80(e))
Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a))
and 1 more
May 3, 2011No Action Indicated (NAI)Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Design plans - Lack of or inadequate (21 CFR 820.30(b))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Documentation (21 CFR 820.75(a))
and 6 more
Sep 17, 2010Voluntary Action Indicated (VAI)Procedures not adequately established or maintained (21 CFR 820.40)
Design plans - Lack of or inadequate (21 CFR 820.30(b))
Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a))
Documentation (21 CFR 820.80(e))
and 7 more
Jan 12, 2010Voluntary Action Indicated (VAI)Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Design plans - Lack of or inadequate (21 CFR 820.30(b))
Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
DHR content (21 CFR 820.184)
and 7 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Apr 9, 2012Class IIZ-2029-2012GMF-RC65 Replacement Surgical Headlight Cable for Cogent Microlink. GMF-RC65…
The GMF-RC65 Replacement Surgical Headlight Cable for Cogent Microlink falls outside of the firm's 510k due to light transmitting fiber…

FDA record score

2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIDec 18, 20196 x 0.251.5
Inspection classified VAIMay 6, 201310 years or older, so it adds no points0
Class II recall Z-2029-2012, this FEI as the recalling firmApr 9, 201210 years or older, so it adds no points0
Inspection classified VAIFeb 3, 201210 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJul 28, 201110 years or older, so it adds no points0
Inspection classified VAISep 17, 201010 years or older, so it adds no points0
Inspection classified VAIJan 12, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing GULF FIBEROPTICS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check GULF FIBEROPTICS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Gulf Fiberoptics, Inc.
Location
OLDSMAR, FL (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.