GULF FIBEROPTICS, INC. FDA Inspection Record
GULF FIBEROPTICS, INC. is an FDA-registered facility (FEI 3003768921) located in OLDSMAR, FL, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Mar 23, 2023: No Action Indicated (NAI)
- Inspections on record
- 8 (5 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jul 28, 2011. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Apr 9, 2012
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 23, 2023 | No Action Indicated (NAI) | No citations listed |
| Dec 18, 2019 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Aug 13, 2015 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) Documentation (21 CFR 820.80(c)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) and 1 more |
| May 6, 2013 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.80(e)) Nonconforming product control (21 CFR 820.90(a)) Lack of or inadequate DHR procedures (21 CFR 820.184) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) |
| Feb 3, 2012 | Voluntary Action Indicated (VAI) | Design input - documentation (21 CFR 820.30(c)) Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) Documentation (21 CFR 820.80(e)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) and 1 more |
| May 3, 2011 | No Action Indicated (NAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design plans - Lack of or inadequate (21 CFR 820.30(b)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Documentation (21 CFR 820.75(a)) and 6 more |
| Sep 17, 2010 | Voluntary Action Indicated (VAI) | Procedures not adequately established or maintained (21 CFR 820.40) Design plans - Lack of or inadequate (21 CFR 820.30(b)) Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) Documentation (21 CFR 820.80(e)) and 7 more |
| Jan 12, 2010 | Voluntary Action Indicated (VAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design plans - Lack of or inadequate (21 CFR 820.30(b)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) DHR content (21 CFR 820.184) and 7 more |
Warning letters
- Jul 28, 2011: Warning letter listed on FDA's Data Dashboard. View compliance record (195372).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 9, 2012 | Class II | Z-2029-2012 | GMF-RC65 Replacement Surgical Headlight Cable for Cogent Microlink.
GMF-RC65… The GMF-RC65 Replacement Surgical Headlight Cable for Cogent Microlink falls outside of the firm's 510k due to light transmitting fiber… |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Dec 18, 2019 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | May 6, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2029-2012, this FEI as the recalling firm | Apr 9, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 3, 2012 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jul 28, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 17, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 12, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing GULF FIBEROPTICS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check GULF FIBEROPTICS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Gulf Fiberoptics, Inc.
- Location
- OLDSMAR, FL (US)
- Source
- device