Johnson & Johnson Surgical Vision Inc FDA Inspection Record
Johnson & Johnson Surgical Vision Inc is an FDA-inspected facility (FEI 3003843509) located in Santa Ana, CA, US. RegDataLab tracks 8 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- May 11, 2016: No Action Indicated (NAI)
- Inspections on record
- 8 (1 OAI, 1 VAI, 6 NAI)
- Last OAI inspection
- Jun 24, 2011
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Form 483 and inspection documents
- 1 in FDA's reading room
- Recalls as the recalling firm
- 12, latest Apr 6, 2022
- Import refusals
- 3
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 11, 2016 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Aug 1, 2013 | No Action Indicated (NAI) | No citations listed |
| Apr 5, 2012 | No Action Indicated (NAI) | No citations listed |
| Oct 20, 2011 | No Action Indicated (NAI) | No citations listed |
| Jun 24, 2011 | Official Action Indicated (OAI) | No citations listed |
| Mar 30, 2010 | Voluntary Action Indicated (VAI) | No citations listed |
| Dec 3, 2009 | No Action Indicated (NAI) | No citations listed |
| Nov 24, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 6, 2022 | Class II | Z-1033-2022 | TRAY RING COVER, REF OM271000, CONTAINS (1) TRAY RING COVER, STERILE R, Rx Only, Potential breach in the sterility barrier for tray ring covers. |
| Aug 6, 2021 | Class II | Z-0028-2022 | TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the… Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal… |
| Jan 27, 2021 | Class II | Z-1163-2021 | TECNIS Toric 1-Piece IOL, Model Number: ZCT150 +23.5D SE 1.50D CYL - Product Usage… Due to the release of nonconforming Intraocular Lenses (IOLs). |
| Dec 3, 2019 | Class II | Z-0867-2020 | Healon GV PRO 0.85 mL, Model (U.S.): 10-2400-14, Model (O.U.S.): 10-2705-14, 10-2715-14… Ophthalmic viscosurgical device may be difficult to remove from the eye, leading to increased postoperative intraocular pressure requiring… |
| Sep 24, 2018 | Class II | Z-1893-2019 | Disposable 25GA Vitrectomy Cutter, REF NGP0025, RX Only, Abbott Medical Optics Inc… Packaging error, 20GA Vitrectomy Cutter may be found in a 25GA package. Use of a Vitrectomy cutter that is a different size than expected… |
| Jun 21, 2018 | Class II | Z-0464-2019 | Abbott TECNIS 1-Piece Aspheric Acrylic IOL with the TECNIS iTEC Preloaded Delivery System… Expired intraocular lenses were distributed. |
| Apr 1, 2017 | Class II | Z-2062-2017 | Healon, Part No. 10290953, 10295210, 10200011, 10200012, 10201012, 10203012, 10213012… and 5 more products Remote possibility that certain solutions in these lots may contain microscopic glass particles due to damage that occurred at the cylinder… |
| Mar 30, 2017 | Class II | Z-2096-2017 | AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and… Packaging error. 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package. |
| Aug 26, 2016 | Class II | Z-2855-2016 | SENSAR Foldable IOL with OptiEdge Design, Model No. AR40e
The product is indicated for… and 9 more products Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the… |
| Nov 6, 2014 | Class II | Z-0530-2015 | Abbott Healon EndoCoat Ophthalmic Viscosurgical Device
3% sodium hyaluronate
0.85 mL 30… AMO has received complaints where the finger grip dislodged while depressing the syringe plunger. The finger grip dislodgement could result… |
Form 483s and inspection documents
FDA record score
9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
14 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1033-2022, this FEI as the recalling firm | Apr 6, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0028-2022, this FEI as the recalling firm | Aug 6, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1163-2021, this FEI as the recalling firm | Jan 27, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0867-2020, this FEI as the recalling firm | Dec 3, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1893-2019, this FEI as the recalling firm | Sep 24, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0464-2019, this FEI as the recalling firm | Jun 21, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2062-2017, this FEI as the recalling firm | Apr 1, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2096-2017, this FEI as the recalling firm | Mar 30, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2855-2016, this FEI as the recalling firm | Aug 26, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0530-2015, this FEI as the recalling firm | Nov 6, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0929-2013, this FEI as the recalling firm | Dec 20, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jun 24, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 30, 2010 | 10 years or older, so it adds no points | 0 |
| Class I recall D-167-6, this FEI as the recalling firm | Nov 18, 2005 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Johnson & Johnson Surgical Vision Inc as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Johnson & Johnson Surgical Vision Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Johnson & Johnson Surgical Vision Inc → Check this supplier
Facility details
- Company
- Johnson & Johnson Surgical Vision Inc
- Location
- Santa Ana, CA (US)
- Source
- inspections