Skeletal Kinetics, Llc FDA Inspection Record

FEI 3003890476 · Inspections, Form 483s, warning letters and import refusals

Skeletal Kinetics, Llc is an FDA-inspected facility (FEI 3003890476) located in Addison, TX, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Oct 5, 2018: No Action Indicated (NAI)
Inspections on record
4 (1 VAI, 3 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
4, latest Apr 6, 2020

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Oct 5, 2018No Action Indicated (NAI)Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Quality Audits - defined intervals (21 CFR 820.22)
Sep 23, 2015No Action Indicated (NAI)Quality audits - auditor independence (21 CFR 820.22)
Jul 9, 2012No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Nov 13, 2009Voluntary Action Indicated (VAI)Procedures for product review,disposition lack of/inadequate (21 CFR 820.90(b)(1))
Report of Death or Serious Injury (21 CFR 803.50(a)(1))
Individual Report of Malfunction (21 CFR 803.50(a)(2))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Apr 6, 2020Class IIZ-2592-2020ACUMED CALLOS INJECT 10CC. Product Number: 65-0010-S, UDI: 813845020245 - Product Usage… and 44 more products
Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required…
Apr 3, 2020Class IIZ-2054-2020Cranio/Sculpt C, 10cc, Bone Void Filler, UDI: 00813845021129 - Product Usage…
Potential product mix-up. The recalling firm has determined that one unit of OsteoVation EX, 3cc that may not have been sterilized was…
Jun 6, 2019Class IIZ-0866-2021OsteoVation, Inject Bone Filler Kits, 5cc (part number 390-2005, manufacturing lot number… and 1 more product
A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No…
May 29, 2015Class IIZ-2209-2015InjectionPlasty 1.0 Bone Void Filler; an injectable and biocompatible calcium phosphate…
Reports of venous uptake, DVT, pulmonary embolism in patients who were implanted with Injection Plasty Bone Void Filler.

FDA record score

3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 17, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

5 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2592-2020, this FEI as the recalling firmApr 6, 20204 x 0.251
Class II recall Z-2054-2020, this FEI as the recalling firmApr 3, 20204 x 0.251
Class II recall Z-0866-2021, this FEI as the recalling firmJun 6, 20194 x 0.251
Class II recall Z-2209-2015, this FEI as the recalling firmMay 29, 201510 years or older, so it adds no points0
Inspection classified VAINov 13, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 17, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 17, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 17, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 17, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 17, 2026Newest record Nov 16, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing Skeletal Kinetics, Llc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Skeletal Kinetics, Llc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Skeletal Kinetics, Llc
Location
Addison, TX (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.