Black Forest Medical GmbH FDA Inspection Record
Black Forest Medical GmbH is an FDA-registered facility (FEI 3003923584) located in FREIBURG Baden-Wurttemberg, DE. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Nov 6, 2014: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (2 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Mar 14, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 6, 2014 | Voluntary Action Indicated (VAI) | Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
| Jun 21, 2012 | Voluntary Action Indicated (VAI) | Procedures not adequately established or maintained (21 CFR 820.40) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Review and evaluation for investigation (21 CFR 820.198(b)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Mar 14, 2025 | Class II | Z-1860-2025 | DORO¿ Easy-Connect Navigation Adaptor; BRAINLAB (item No. 1204.002) Potential of compromised compatibility resulting in influence on the positioning of the reference array and an influence of the function of… |
| Jul 5, 2018 | Class I | Z-2970-2018 | DORO LUCENT Base Unit, REF 1101.021 Headrest System used in open and percutaneous… and 2 more products The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one… |
| Nov 2, 2017 | Class II | Z-0193-2018 | pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per… The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at… |
FDA record score
9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1860-2025, this FEI as the recalling firm | Mar 14, 2025 | 4 x 1 | 4 |
| Class I recall Z-2970-2018, this FEI as the recalling firm | Jul 5, 2018 | 15 x 0.25 | 3.8 |
| Class II recall Z-0193-2018, this FEI as the recalling firm | Nov 2, 2017 | 4 x 0.25 | 1 |
| Inspection classified VAI | Nov 6, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 21, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Black Forest Medical GmbH as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Black Forest Medical GmbH yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Pro-Med Instruments Gmbh
- Location
- FREIBURG Baden-Wurttemberg (DE)
- Source