Cellestis Inc FDA Inspection Record
Cellestis Inc is an FDA-inspected facility (FEI 3003964343) located in Santa Clarita, CA, US. RegDataLab tracks 3 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Mar 25, 2015: Official Action Indicated (OAI)
- Inspections on record
- 3 (1 OAI, 1 VAI, 1 NAI)
- Last OAI inspection
- Mar 25, 2015
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Aug 26, 2015. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Sep 27, 2012
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 25, 2015 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) DMR - not or inadequately maintained (21 CFR 820.181) Records not kept (21 CFR 806.20(a)) and 3 more |
| Sep 15, 2011 | No Action Indicated (NAI) | No citations listed |
| Oct 6, 2008 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) |
Warning letters
- Aug 26, 2015: Warning letter listed on FDA's Data Dashboard. View compliance record (458031).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 27, 2012 | Class II | Z-0888-2013 | VACUETTE Cellestis 1 ml QuantiFERON¿ - TB ANTIGEN; Shake 10x; REF 454074; HE0105; 13X75 50… and 1 more product Cellestis is recalling QuantiFERON¿¿ - TB Gold TB Blood Collection Tubes and HA TB Blood Collection Tubes due to potential contamination. |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Warning letter on FDA's dashboard, which does not show close-outs | Aug 26, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Mar 25, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0888-2013, this FEI as the recalling firm | Sep 27, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 6, 2008 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Cellestis Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.