Organ Recovery Systems, Inc. FDA Inspection Record

FEI 3004068499 · Inspections, Form 483s, warning letters and import refusals

Organ Recovery Systems, Inc. is an FDA-registered facility (FEI 3004068499) located in Itasca, IL, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Oct 16, 2018: No Action Indicated (NAI)
Inspections on record
5 (1 OAI, 2 VAI, 2 NAI)
Last OAI inspection
Feb 10, 2017
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Apr 20, 2017. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
2, latest Jan 30, 2020

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Oct 16, 2018No Action Indicated (NAI)No citations listed
Feb 10, 2017Official Action Indicated (OAI)Lack of or inadequate procedures for distribution (21 CFR 820.160(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Design validation - documentation (21 CFR 820.30(g))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
and 1 more
Apr 7, 2015No Action Indicated (NAI)Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Lack of or inadequate procedures (21 CFR 820.100(a))
Documentation (21 CFR 820.100(b))
and 1 more
Nov 8, 2011Voluntary Action Indicated (VAI)No citations listed
Dec 9, 2009Voluntary Action Indicated (VAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jan 30, 2020Class IIZ-1437-2020LKT200 Perfusion Circuit, part of Lifeport Kidney Perfusion Transporter KTR - Product…
Observed leakage at the sample port assembly which could potentially lead to loss of perfusion solution needed to maintain continuous…
Dec 14, 2016Class IIZ-1111-2017SPS-1, Static Preservation Solution [1 or 2 liter bags], Sterile; Organ Recovery Systems…
Organ preservation fluid recalled due to potential for bacterial contamination.

FDA record score

17 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 28, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

6 records behind this number
RecordDatePoints x age weightAdds
Warning letter on FDA's dashboard, which does not show close-outsApr 20, 201730 x 0.257.5
Inspection classified OAIFeb 10, 201730 x 0.257.5
Class II recall Z-1437-2020, this FEI as the recalling firmJan 30, 20204 x 0.251
Class II recall Z-1111-2017, this FEI as the recalling firmDec 14, 20164 x 0.251
Inspection classified VAINov 8, 201110 years or older, so it adds no points0
Inspection classified VAIDec 9, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 28, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 28, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 28, 2026Newest record Nov 19, 2026Current
Warning letter textNov 28, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 28, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Organ Recovery Systems, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Organ Recovery Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Organ Recovery Systems, Inc. → Check this supplier

Facility details

FDA Legal Name
ORGAN RECOVERY SYSTEMS, INC.
Company
Organ Recovery Systems, Inc.
Location
Itasca, IL (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.