OPTI MEDICAL SYSTEMS INC. FDA Inspection Record
OPTI MEDICAL SYSTEMS INC. is an FDA-registered facility (FEI 3004102403) located in ROSWELL, GA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 29, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (1 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Jun 25, 2021
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 29, 2017 | Voluntary Action Indicated (VAI) | Periodic equipment inspections (21 CFR 820.70(g)(2)) Lack of or inadequate procedures (21 CFR 820.100(a)) Rationale documented for no investigation (21 CFR 820.198(b)) Supplier notification of changes (21 CFR 820.50(b)) and 3 more |
| Jun 28, 2013 | No Action Indicated (NAI) | No citations listed |
| Nov 19, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 25, 2021 | Class II | Z-2310-2021 | OPTI SARS-CoV-2 RT-PCR Test, REF Numbers:
a) 99-57003
b) 99-57004 The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected… |
| Aug 9, 2013 | Class II | Z-0029-2014 | Model BP7507 OPTI LION E-Plus Cassettes 25 Cassettes with Sample Probes… OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more… |
| Jul 3, 2012 | Class II | Z-2440-2012 | BP7561 OPTI CCA-TS B-Lac Cassette (25 per box)
Cassettes are consumables used in the… The values reported for PCO2 by BP7561 B-Lac cassettes are biased at high concentrations when used at altitudes above 2800 feet (853… |
FDA record score
3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Mar 29, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-2310-2021, this FEI as the recalling firm | Jun 25, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0029-2014, this FEI as the recalling firm | Aug 9, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2440-2012, this FEI as the recalling firm | Jul 3, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing OPTI MEDICAL SYSTEMS INC. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for OPTI MEDICAL SYSTEMS INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about OPTI MEDICAL SYSTEMS INC. → Check this supplier
Facility details
- Company
- OPTI Medical Systems, Inc
- Location
- ROSWELL, GA (US)
- Source
- device