Solta Medical, Inc. FDA Inspection Record
Solta Medical, Inc. is an FDA-registered facility (FEI 3004106598) located in Bothell, WA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 15, 2016: No Action Indicated (NAI)
- Inspections on record
- 3 (1 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Feb 16, 2023
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 15, 2016 | No Action Indicated (NAI) | Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) |
| Jan 23, 2012 | No Action Indicated (NAI) | Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Mar 4, 2010 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 16, 2023 | Class II | Z-1354-2023 | thermage THERMAGE CPT SYSTEM REF TG-2B *** Solta Medical, Inc.
thermage RF GENERATOR REF… A service unit was potentially improperly tested and calibrated during manufacturing on a failed test station. |
| Dec 30, 2019 | Class II | Z-1681-2020 | Thermage Face Tip 3.0, REF/UDI: TTNS3,00C3-400/UDI (01)000850608002377; REF… Complaints of sparking of the tips of the electrosurgical device used in RF surgical procedures. The use of the affected treatment tips… |
| Jun 1, 2017 | Class II | Z-3047-2017 | VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA
The… and 1 more product Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts… |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1354-2023, this FEI as the recalling firm | Feb 16, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1681-2020, this FEI as the recalling firm | Dec 30, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-3047-2017, this FEI as the recalling firm | Jun 1, 2017 | 4 x 0.25 | 1 |
| Inspection classified VAI | Mar 4, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Solta Medical, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Solta Medical, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Solta Medical Inc
- Location
- Bothell, WA (US)
- Source
- device