HEARTSINE TECHNOLOGIES, LTD FDA Inspection Record

FEI 3004123209 · Inspections, Form 483s, warning letters and import refusals

HEARTSINE TECHNOLOGIES, LTD is an FDA-registered facility (FEI 3004123209) located in BELFAST Northern Ireland, GB. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Apr 12, 2018: No Action Indicated (NAI)
Inspections on record
3 (1 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
10, latest Jul 27, 2026
Import refusals
4

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Apr 12, 2018No Action Indicated (NAI)No citations listed
Feb 12, 2015No Action Indicated (NAI)No citations listed
Jan 28, 2011Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 27, 2026Class IIZ-3120-2026HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI…
Due to a damaged (over stretched) membrane tail that causes an intermittent fault.
Jul 27, 2026Class IIIZ-3270-2026HeartSine samaritan PAD (Public Access Defibrillator); HeartSine SAM 350P 350P US…
Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the…
Sep 18, 2025Class IIZ-0288-2026HeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07…
Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.
Jun 30, 2025Class IIZ-2174-2025HeartSine SAM 350P, HeartSine SAM 360P, and HeartSine SAM 450P The HeartSine…
Due to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks)
Apr 8, 2024Class IIZ-1820-2024HeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P
Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to…
Feb 14, 2024Class IIZ-1141-2024HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM…
Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR…
Apr 26, 2023Class IIZ-2143-2023HeartSine, REF: PAD-PAK-01, PAD-PAK-03 AND PAD-PAK-03j, CE0123 sold separately and with…
Single use battery and electrode cartridge may be rendered inoperable due to depleted battery cells which could prevent the defibrillator…
Feb 1, 2023Class IIZ-1217-2023Heartsine Samaritan PAD: SAM 350P semi-automatic defibrillator, SAM 360P fully automatic…
Automated external defibrillators were potentially distributed with the incorrect language configuration, which can lead to a delay in…
Jun 12, 2014Class IIZ-2115-2014Pad-Pak and Pedi-Pak accessories for the HeartSine Samaritan PAD 300P and 350P device (SAM…
A small number of sealed foil pouches containing the electrodes were found to be difficult to open, resulting in potential delay in therapy.
Sep 13, 2012Class IZ-0124-2013HeartSine Samaritan(R) 300/300P PAD devices with serial numbers 0400000501 to 0700032917…
Certain Samaritan 300/300P PAD devices may experience one or both of the following conditions that could affect the ability to deliver…

FDA record score

22 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 14, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-3120-2026, this FEI as the recalling firmJul 27, 20264 x 14
Class II recall Z-0288-2026, this FEI as the recalling firmSep 18, 20254 x 14
Class II recall Z-2174-2025, this FEI as the recalling firmJun 30, 20254 x 14
Class II recall Z-1820-2024, this FEI as the recalling firmApr 8, 20244 x 0.62.4
Class II recall Z-1141-2024, this FEI as the recalling firmFeb 14, 20244 x 0.62.4
Class II recall Z-2143-2023, this FEI as the recalling firmApr 26, 20234 x 0.62.4
Class II recall Z-1217-2023, this FEI as the recalling firmFeb 1, 20234 x 0.62.4
Class II recall Z-2115-2014, this FEI as the recalling firmJun 12, 201410 years or older, so it adds no points0
Class I recall Z-0124-2013, this FEI as the recalling firmSep 13, 201210 years or older, so it adds no points0
Inspection classified VAIJan 28, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 14, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 14, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 14, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 14, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 2, 2026Newest record Oct 22, 2026Current

Reviewing HEARTSINE TECHNOLOGIES, LTD as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check HEARTSINE TECHNOLOGIES, LTD yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
HeartSine Technologies Ltd
Location
BELFAST Northern Ireland (GB)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.