LUMENIS, LTD. FDA Inspection Record
LUMENIS, LTD. is an FDA-registered facility (FEI 3004135191) located in YOKNEAM Northern, IL. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 2, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 7, latest Jul 22, 2025
- Import refusals
- 7
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 2, 2017 | Voluntary Action Indicated (VAI) | Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Design plans- updated (21 CFR 820.30(b)) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) and 8 more |
| May 28, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 22, 2025 | Class II | Z-2344-2025 | OtoLase Starter Kit;
UPN: AC-2003748.
Indicated for a
variety of surgical uses… The potential for unsterilized product within finished product labeled as sterile. |
| Jan 23, 2025 | Class II | Z-1290-2025 | (1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System;
(2) Lumenis Pulse 120H… The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller… |
| Sep 25, 2016 | Class II | Z-0301-2017 | FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip… The cleaning instructions provided in FemTouch Operator's Manual might be ineffective for a fully assembled device and may lead to a… |
| Nov 17, 2015 | Class II | Z-1669-2016 | Lumenis M22 System Model Number: GA-0005400 (M22 IPL + YAG Module) with Acne Filter… Lumenis Ltd Announces a Field Action of the M22 IPL Acne Filters for the Lumenis M22 IPL Hand Piece due to the Risk of Superficial Burns… |
| Jul 30, 2015 | Class II | Z-2770-2015 | Lumenis VersaCut+ Tissue Morcellator GA-0007500
The Lumenis VersaCut / VersaCut+ Tissue… There is a probability the hand piece may operate inconsistently and unpredictably during the morcellation procedure resulting in damage to… |
| Mar 25, 2015 | Class II | Z-1519-2015 | Light Sheer Desire Diode Laser System with XC Handpiece Accessory options. Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended… |
| Jan 2, 2013 | Class I | Z-0795-2013 | The Lumenis VersaCut Tissue Morcellator
Product Usage: is intended for use under… Lumenis Ltd is recalling certain models of VersaCut Morcellator devices to perform corrective labeling as a result of three reported events… |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2344-2025, this FEI as the recalling firm | Jul 22, 2025 | 4 x 1 | 4 |
| Class II recall Z-1290-2025, this FEI as the recalling firm | Jan 23, 2025 | 4 x 1 | 4 |
| Inspection classified VAI | Feb 2, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0301-2017, this FEI as the recalling firm | Sep 25, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1669-2016, this FEI as the recalling firm | Nov 17, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2770-2015, this FEI as the recalling firm | Jul 30, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1519-2015, this FEI as the recalling firm | Mar 25, 2015 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0795-2013, this FEI as the recalling firm | Jan 2, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing LUMENIS, LTD. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check LUMENIS, LTD. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- LUMENIS, LTD.
- Location
- YOKNEAM Northern (IL)
- Source
- device