LUMENIS, LTD. FDA Inspection Record

FEI 3004135191 · Inspections, Form 483s, warning letters and import refusals

LUMENIS, LTD. is an FDA-registered facility (FEI 3004135191) located in YOKNEAM Northern, IL. RegDataLab tracks 2 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Feb 2, 2017: Voluntary Action Indicated (VAI)
Inspections on record
2 (1 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
7, latest Jul 22, 2025
Import refusals
7

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Feb 2, 2017Voluntary Action Indicated (VAI)Design input - Lack of or inadequate procedures (21 CFR 820.30(c))
Design plans- updated (21 CFR 820.30(b))
Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
and 8 more
May 28, 2009No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 22, 2025Class IIZ-2344-2025OtoLase Starter Kit; UPN: AC-2003748. Indicated for a variety of surgical uses…
The potential for unsterilized product within finished product labeled as sterile.
Jan 23, 2025Class IIZ-1290-2025(1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H…
The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller…
Sep 25, 2016Class IIZ-0301-2017FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip…
The cleaning instructions provided in FemTouch Operator's Manual might be ineffective for a fully assembled device and may lead to a…
Nov 17, 2015Class IIZ-1669-2016Lumenis M22 System Model Number: GA-0005400 (M22 IPL + YAG Module) with Acne Filter…
Lumenis Ltd Announces a Field Action of the M22 IPL Acne Filters for the Lumenis M22 IPL Hand Piece due to the Risk of Superficial Burns…
Jul 30, 2015Class IIZ-2770-2015Lumenis VersaCut+ Tissue Morcellator GA-0007500 The Lumenis VersaCut / VersaCut+ Tissue…
There is a probability the hand piece may operate inconsistently and unpredictably during the morcellation procedure resulting in damage to…
Mar 25, 2015Class IIZ-1519-2015Light Sheer Desire Diode Laser System with XC Handpiece Accessory options.
Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended…
Jan 2, 2013Class IZ-0795-2013The Lumenis VersaCut Tissue Morcellator Product Usage: is intended for use under…
Lumenis Ltd is recalling certain models of VersaCut Morcellator devices to perform corrective labeling as a result of three reported events…

FDA record score

10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

8 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2344-2025, this FEI as the recalling firmJul 22, 20254 x 14
Class II recall Z-1290-2025, this FEI as the recalling firmJan 23, 20254 x 14
Inspection classified VAIFeb 2, 20176 x 0.251.5
Class II recall Z-0301-2017, this FEI as the recalling firmSep 25, 201610 years or older, so it adds no points0
Class II recall Z-1669-2016, this FEI as the recalling firmNov 17, 201510 years or older, so it adds no points0
Class II recall Z-2770-2015, this FEI as the recalling firmJul 30, 201510 years or older, so it adds no points0
Class II recall Z-1519-2015, this FEI as the recalling firmMar 25, 201510 years or older, so it adds no points0
Class I recall Z-0795-2013, this FEI as the recalling firmJan 2, 201310 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing LUMENIS, LTD. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check LUMENIS, LTD. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
LUMENIS, LTD.
Location
YOKNEAM Northern (IL)
Source
device