BIOCARE MEDICAL, LLC FDA Inspection Record
BIOCARE MEDICAL, LLC is an FDA-registered facility (FEI 3004140393) located in Pacheco, CA, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 11, 2016: No Action Indicated (NAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 10, latest Sep 25, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 11, 2016 | No Action Indicated (NAI) | No citations listed |
| Apr 27, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate organizational structure (21 CFR 820.20(b)) Report of risk to health (21 CFR 806.10(a)(1)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 25, 2024 | Class II | Z-0315-2025 | intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB
Buffer, Catalog… Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may… |
| Feb 25, 2021 | Class II | Z-1333-2021 | BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model… Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction… |
| Oct 10, 2019 | Class II | Z-2291-2020 | 4plus Streptavidin HRP Label, Conjugated Streptavidin Horseradish Peroxidase, Detection… Firm has identified product has no staining. If used, may result in invalid test results when used with controls in the clinical setting… |
| Apr 24, 2018 | Class II | Z-2360-2018 | Vulcan Fast Red Chromogen Kit 2, part of an immunohistochemistry (IHC) procedure using an… Chromogen failed to develop due to a stability issue caused by environmental factors with a raw material in the Vulcan Fast Red Buffer. |
| Mar 15, 2018 | Class II | Z-3198-2018 | URO-3 Triple Stain (CD44 + p53) with CK20, Catalog #: PM370TSAA, Package Type: Bag, which… Immunohistochemistry device kits found to display incorrect staining pattern due to mislabeling of two components. |
| Mar 31, 2016 | Class III | Z-1523-2016 | Mach 4 Mouse Probe vial within the Mach 4 Universal HRP-Polymer Kit;
Catalog number… Biocare Medical has identified a single lot (082715) of its Mach 4 HRP Polymer kit in which some few Mach 4 Mouse Probe vials may have been… |
| Sep 8, 2015 | Class II | Z-0252-2016 | Smooth Muscle Action (SMA) Immunohistochemistry reagent. Catalog numbers: CME305A-0.1ml… Certain catalog numbers and lots of Smooth Muscle Actin may stain some cell types that are typically expected to be negative for SMA. |
| May 6, 2015 | Class III | Z-1676-2015 | D2-40 (Lymphatic Marker); For In Vitro Diagnostic;
Catalog number: PM266AA
Product… Firm discovered that incorrect Instructions for Use were shipped with one lot of D2-40. |
| Mar 17, 2015 | Class II | Z-1535-2015 | CD99
Catalog number PM008 AA;
For In Vitro Diagnostic Use.
Product Usage:
Provides… A drop in staining intensity over time has been observed. These lots may not meet their labeled shelf life. |
| Dec 12, 2014 | Class II | Z-0965-2015 | MUM-1 (Multiple myeloma oncogene-1), catalog number PRM352 AA;
Product Usage: For In… Three lots of MUM-1 may not meet their labeled shelf life, as a drop in staining intensity over time has been observed. |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0315-2025, this FEI as the recalling firm | Sep 25, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1333-2021, this FEI as the recalling firm | Feb 25, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2291-2020, this FEI as the recalling firm | Oct 10, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2360-2018, this FEI as the recalling firm | Apr 24, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-3198-2018, this FEI as the recalling firm | Mar 15, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0252-2016, this FEI as the recalling firm | Sep 8, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 27, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1535-2015, this FEI as the recalling firm | Mar 17, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0965-2015, this FEI as the recalling firm | Dec 12, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 17, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 17, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 17, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 17, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 17, 2026 | Newest record Nov 16, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing BIOCARE MEDICAL, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check BIOCARE MEDICAL, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Biocare Medical, LLC
- Location
- Pacheco, CA (US)
- Source
- device