CENORIN, LLC FDA Inspection Record

FEI 3004153171 · Inspections, Form 483s, warning letters and import refusals

CENORIN, LLC is an FDA-registered facility (FEI 3004153171) located in KENT, WA, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Feb 14, 2020: No Action Indicated (NAI)
Inspections on record
5 (1 OAI, 1 VAI, 3 NAI)
Last OAI inspection
May 2, 2014
Warning letters
1 listed by FDA for this FEI, latest Feb 12, 2015
Recalls as the recalling firm
4, latest Apr 9, 2018

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Feb 14, 2020No Action Indicated (NAI)No citations listed
Dec 23, 2015Voluntary Action Indicated (VAI)Design input - documentation (21 CFR 820.30(c))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Lack of or inadequate procedures (21 CFR 820.100(a))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
May 2, 2014Official Action Indicated (OAI)Design input - documentation (21 CFR 820.30(c))
Lack of or inadequate instructions (21 CFR 820.170(a))
Installer records (21 CFR 820.170(b))
Documentation (21 CFR 820.100(b))
Jun 20, 2012No Action Indicated (NAI)No citations listed
Oct 26, 2009No Action Indicated (NAI)Documentation (21 CFR 820.100(b))
Quality Audits - defined intervals (21 CFR 820.22)

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Apr 9, 2018Class IIZ-2288-2018HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160…
There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.
Jun 7, 2017Class IIZ-3032-2017ThermaSure(TM), a CENORIN(TM) product, SERIES 130 MEDICAL DEVICE DRYER: (a) Model 135; (b)… and 2 more products
Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.
Apr 24, 2014Class IIZ-1729-2014HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610…
Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting…
Jan 24, 2013Class IIZ-1995-2014HLD System, Model 610, Medical Device Cleaning and High Level Disinfection…
The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain…

FDA record score

2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2288-2018, this FEI as the recalling firmApr 9, 20184 x 0.251
Class II recall Z-3032-2017, this FEI as the recalling firmJun 7, 20174 x 0.251
Inspection classified VAIDec 23, 201510 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsFeb 12, 201510 years or older, so it adds no points0
Inspection classified OAIMay 2, 201410 years or older, so it adds no points0
Class II recall Z-1729-2014, this FEI as the recalling firmApr 24, 201410 years or older, so it adds no points0
Class II recall Z-1995-2014, this FEI as the recalling firmJan 24, 201310 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing CENORIN, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check CENORIN, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Cenorin, LLC
Location
KENT, WA (US)
Source
device