Integra LifeSciences Braintree Inc. FDA Inspection Record

FEI 3004170064 · Inspections, Form 483s, warning letters and import refusals

Integra LifeSciences Braintree Inc. is an FDA-registered facility (FEI 3004170064) located in Braintree, MA, US. RegDataLab tracks 6 FDA inspections and 3 compliance actions at this facility.

What FDA found at this site

Latest inspection
May 17, 2023: Voluntary Action Indicated (VAI)
Inspections on record
6 (1 OAI, 3 VAI, 2 NAI)
Last OAI inspection
Nov 2, 2018
Warning letters
3 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jul 17, 2023. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
2, latest May 23, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
May 17, 2023Voluntary Action Indicated (VAI)Lack of or inadequate procedures for storage. (21 CFR 820.150)
Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e))
Software validation for automated processes (21 CFR 820.70(i))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 3 more
Nov 12, 2021Voluntary Action Indicated (VAI)Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b))
Examination for accuracy (21 CFR 820.120(b))
Environmental control Lack of or inadequate procedures (21 CFR 820.70(c))
Lack of or inadequate process validation (21 CFR 820.75(a))
and 2 more
Nov 2, 2018Official Action Indicated (OAI)No citations listed
Jun 1, 2015Voluntary Action Indicated (VAI)Documentation (21 CFR 820.100(b))
Feb 15, 2012No Action Indicated (NAI)No citations listed
Jan 15, 2009No Action Indicated (NAI)No citations listed

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 23, 2023Class IIZ-2069-2023SurgiMend; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 1.0 5x6 cm, 1.0… and 16 more products
Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing.
Jun 15, 2017Class IIZ-0383-2018PriMatrix¿ is an acellular dermal tissue matrix derived from fetal bovine dermis. The…
During Final Assembly Operation, it was noticed that some of the PriMatrix Instructions for Use (IFUs) from Lot RM1608011 Part 520-002v10…

FDA record score

42 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

9 records behind this number
RecordDatePoints x age weightAdds
Warning letter, no close-out letter listedJul 17, 202330 x 0.618
Warning letter, no close-out letter listedMar 6, 201930 x 0.257.5
Inspection classified OAINov 2, 201830 x 0.257.5
Inspection classified VAIMay 17, 20236 x 0.63.6
Class II recall Z-2069-2023, this FEI as the recalling firmMay 23, 20234 x 0.62.4
Inspection classified VAINov 12, 20216 x 0.251.5
Class II recall Z-0383-2018, this FEI as the recalling firmJun 15, 20174 x 0.251
Inspection classified VAIJun 1, 201510 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsMay 29, 201510 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Integra LifeSciences Braintree Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Integra LifeSciences Braintree Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
TEI Biosciences, Inc.
Location
Braintree, MA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.