Integra LifeSciences Braintree Inc. FDA Inspection Record
Integra LifeSciences Braintree Inc. is an FDA-registered facility (FEI 3004170064) located in Braintree, MA, US. RegDataLab tracks 6 FDA inspections and 3 compliance actions at this facility.
What FDA found at this site
- Latest inspection
- May 17, 2023: Voluntary Action Indicated (VAI)
- Inspections on record
- 6 (1 OAI, 3 VAI, 2 NAI)
- Last OAI inspection
- Nov 2, 2018
- Warning letters
- 3 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jul 17, 2023. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest May 23, 2023
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 17, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures for storage. (21 CFR 820.150) Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e)) Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 3 more |
| Nov 12, 2021 | Voluntary Action Indicated (VAI) | Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b)) Examination for accuracy (21 CFR 820.120(b)) Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) Lack of or inadequate process validation (21 CFR 820.75(a)) and 2 more |
| Nov 2, 2018 | Official Action Indicated (OAI) | No citations listed |
| Jun 1, 2015 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) |
| Feb 15, 2012 | No Action Indicated (NAI) | No citations listed |
| Jan 15, 2009 | No Action Indicated (NAI) | No citations listed |
Warning letters
- Jul 17, 2023: CGMP/QSR/Medical Devices/Adulterated. Read the letter on FDA.gov.
- Mar 6, 2019. Read the letter on FDA.gov.
- May 29, 2015: Warning letter listed on FDA's Data Dashboard. View compliance record (459703).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 23, 2023 | Class II | Z-2069-2023 | SurgiMend; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 1.0 5x6 cm, 1.0… and 16 more products Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing. |
| Jun 15, 2017 | Class II | Z-0383-2018 | PriMatrix¿ is an acellular dermal tissue matrix derived from fetal bovine dermis. The… During Final Assembly Operation, it was noticed that some of the PriMatrix Instructions for Use (IFUs) from Lot RM1608011 Part 520-002v10… |
FDA record score
42 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Warning letter, no close-out letter listed | Jul 17, 2023 | 30 x 0.6 | 18 |
| Warning letter, no close-out letter listed | Mar 6, 2019 | 30 x 0.25 | 7.5 |
| Inspection classified OAI | Nov 2, 2018 | 30 x 0.25 | 7.5 |
| Inspection classified VAI | May 17, 2023 | 6 x 0.6 | 3.6 |
| Class II recall Z-2069-2023, this FEI as the recalling firm | May 23, 2023 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Nov 12, 2021 | 6 x 0.25 | 1.5 |
| Class II recall Z-0383-2018, this FEI as the recalling firm | Jun 15, 2017 | 4 x 0.25 | 1 |
| Inspection classified VAI | Jun 1, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | May 29, 2015 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |