Ra Medical Systems, Inc. FDA Inspection Record

FEI 3004217728 · Inspections, Form 483s, warning letters and import refusals

Ra Medical Systems, Inc. is an FDA-inspected facility (FEI 3004217728) located in Carlsbad, CA, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Dec 11, 2019: No Action Indicated (NAI)
Inspections on record
5 (3 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
7, latest Oct 16, 2020

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Dec 11, 2019No Action Indicated (NAI)Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
DMR - not or inadequately maintained (21 CFR 820.181)
Lack of procedures, or not maintained (21 CFR 820.40)
May 24, 2018Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Investigation of device failures (21 CFR 820.198(c))
Document locations, Dissemination, etc. (21 CFR 820.40(a))
and 1 more
Nov 5, 2014No Action Indicated (NAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2))
Production and Process Change Procedures, lack of or Inad. (21 CFR 820.70(b))
Nov 1, 2012Voluntary Action Indicated (VAI)No citations listed
May 24, 2010Voluntary Action Indicated (VAI)DHR content (21 CFR 820.184)
Not approved or obsolete document retrieval (21 CFR 820.40(a))
Environmental control Lack of or inadequate procedures (21 CFR 820.70(c))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Oct 16, 2020Class IIZ-0569-2021DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in…
Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on…
Jul 24, 2020Class IIZ-0050-2021DABRA RA-308 Excimer Laser, SN: ******* - Product Usage: The device is intended for use in…
Due to the taller wheel mounting on the laser, the laser may fall over potentially causing user or patient injury.
Jan 23, 2020Class IIZ-2497-2020DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer laser…
A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that…
Sep 19, 2019Class IIZ-2006-2020Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 18-090400-094, USE…
The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to…
Aug 23, 2019Class IIZ-1631-2020Pharos Excimer Laser, Model: EX-308 - Product Usage: The Pharos EX-308 Excimer Laser…
At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device…
Feb 15, 2018Class IIZ-2218-2019The DABRA Laser (model RA-308) is designed to provide ultraviolet laser energy…
Lasers/Catheters did not calibrate during set-up prior to use.
Jun 30, 2014Class IIZ-2123-2014PHAROS Excimer Laser, Model No. EX-308 UVB phototherapy for the treatment of psoriasis…
RA Medical Systems is recalling the PHAROS EX-308 Excimer Laser because there is a possibility that the laser may calibrate with the iris…

FDA record score

8 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIMay 24, 20186 x 0.251.5
Class II recall Z-0569-2021, this FEI as the recalling firmOct 16, 20204 x 0.251
Class II recall Z-0050-2021, this FEI as the recalling firmJul 24, 20204 x 0.251
Class II recall Z-2497-2020, this FEI as the recalling firmJan 23, 20204 x 0.251
Class II recall Z-2006-2020, this FEI as the recalling firmSep 19, 20194 x 0.251
Class II recall Z-1631-2020, this FEI as the recalling firmAug 23, 20194 x 0.251
Class II recall Z-2218-2019, this FEI as the recalling firmFeb 15, 20184 x 0.251
Class II recall Z-2123-2014, this FEI as the recalling firmJun 30, 201410 years or older, so it adds no points0
Inspection classified VAINov 1, 201210 years or older, so it adds no points0
Inspection classified VAIMay 24, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Ra Medical Systems, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ra Medical Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ra Medical Systems, Inc. → Check this supplier

Facility details

Company
Ra Medical Systems, Inc.
Location
Carlsbad, CA (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.