Ra Medical Systems, Inc. FDA Inspection Record
Ra Medical Systems, Inc. is an FDA-inspected facility (FEI 3004217728) located in Carlsbad, CA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Dec 11, 2019: No Action Indicated (NAI)
- Inspections on record
- 5 (3 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 7, latest Oct 16, 2020
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Dec 11, 2019 | No Action Indicated (NAI) | Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) DMR - not or inadequately maintained (21 CFR 820.181) Lack of procedures, or not maintained (21 CFR 820.40) |
| May 24, 2018 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Document locations, Dissemination, etc. (21 CFR 820.40(a)) and 1 more |
| Nov 5, 2014 | No Action Indicated (NAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2)) Production and Process Change Procedures, lack of or Inad. (21 CFR 820.70(b)) |
| Nov 1, 2012 | Voluntary Action Indicated (VAI) | No citations listed |
| May 24, 2010 | Voluntary Action Indicated (VAI) | DHR content (21 CFR 820.184) Not approved or obsolete document retrieval (21 CFR 820.40(a)) Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 16, 2020 | Class II | Z-0569-2021 | DABRA Laser (RA-308 Excimer Laser)
Intended Use: For use in ablating a channel in… Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on… |
| Jul 24, 2020 | Class II | Z-0050-2021 | DABRA RA-308 Excimer Laser, SN: ******* - Product Usage: The device is intended for use in… Due to the taller wheel mounting on the laser, the laser may fall over potentially causing user or patient injury. |
| Jan 23, 2020 | Class II | Z-2497-2020 | DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer laser… A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that… |
| Sep 19, 2019 | Class II | Z-2006-2020 | Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 18-090400-094, USE… The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to… |
| Aug 23, 2019 | Class II | Z-1631-2020 | Pharos Excimer Laser, Model: EX-308 - Product Usage: The Pharos EX-308 Excimer Laser… At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device… |
| Feb 15, 2018 | Class II | Z-2218-2019 | The DABRA Laser (model RA-308) is designed to provide ultraviolet laser energy… Lasers/Catheters did not calibrate during set-up prior to use. |
| Jun 30, 2014 | Class II | Z-2123-2014 | PHAROS Excimer Laser, Model No. EX-308
UVB phototherapy for the treatment of psoriasis… RA Medical Systems is recalling the PHAROS EX-308 Excimer Laser because there is a possibility that the laser may calibrate with the iris… |
FDA record score
8 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | May 24, 2018 | 6 x 0.25 | 1.5 |
| Class II recall Z-0569-2021, this FEI as the recalling firm | Oct 16, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0050-2021, this FEI as the recalling firm | Jul 24, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2497-2020, this FEI as the recalling firm | Jan 23, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2006-2020, this FEI as the recalling firm | Sep 19, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1631-2020, this FEI as the recalling firm | Aug 23, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2218-2019, this FEI as the recalling firm | Feb 15, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2123-2014, this FEI as the recalling firm | Jun 30, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 1, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 24, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Ra Medical Systems, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Ra Medical Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Ra Medical Systems, Inc. → Check this supplier
Facility details
- Company
- Ra Medical Systems, Inc.
- Location
- Carlsbad, CA (US)
- Source
- inspections