CEPHEID FDA Inspection Record
CEPHEID is an FDA-registered facility (FEI 3004530258) located in Sunnyvale, CA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 9, 2017: No Action Indicated (NAI)
- Inspections on record
- 5 (3 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 9, latest Nov 5, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 9, 2017 | No Action Indicated (NAI) | No citations listed |
| Jun 16, 2015 | Voluntary Action Indicated (VAI) | Investigation of device failures (21 CFR 820.198(c)) Design validation - user needs and intended uses (21 CFR 820.30(g)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Jan 13, 2012 | Voluntary Action Indicated (VAI) | Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
| Oct 29, 2010 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Complaints (21 CFR 820.198(a)) Lack of System for Determining MDR Events (21 CFR 803.17(a)(2)) |
| Sep 16, 2009 | No Action Indicated (NAI) | Storage procedures to prevent mix-ups (21 CFR 820.150(a)) Storage procedures to avoid release of unsuitable product (21 CFR 820.150(a)) Control/distribution procedures (21 CFR 820.160(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 5, 2025 | Class II | Z-0724-2026 | Xpert MTB/RIF. Model Number: GXMTB/RIF-US-10. Product testing did not meet expected stability criteria. |
| Oct 30, 2025 | Class II | Z-0959-2026 | Brand Name: Xpert¿ BCR-ABL Ultra
Model/Catalog Number: GXBCRABL-US-10 and 1 more product Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results. |
| Aug 6, 2025 | Class II | Z-0414-2026 | Xpert SA Nasal Complete
Catalog Number: GXSACOMP-10
UDI-DI code: 07332940000158
Xpert… As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type. |
| Jun 20, 2024 | Class II | Z-2584-2024 | Cepheid, Sample Collection Device, Part: 900-0370 Specimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard… |
| Apr 24, 2023 | Class II | Z-0215-2024 | Xpert Xpress Strep A, XPRSTREPA-10, P/N: 302-2294 IVD, Rx Only and 1 more product pipettes packaged in Strep A test may fail to dispense an adequate sample volume which could result in "invalid" results or no result test… |
| Oct 9, 2017 | Class II | Z-0592-2018 | Xpert CT/NG Catalog GXCT/NG-120 Batch 1000049568, Catalog GXCT/NG-10 Batch 1000053075… Cepheid has received reports that some of our customers have experienced a higher number of temperature drift errors than would typically… |
| Sep 15, 2017 | Class II | Z-0028-2018 | Xpert EV Assay, Catalog GXEV-100N-10.
The Cepheid Xpert EV assay is a reverse… Customers have reported a higher than expected number of invalid test results when using the recalled product per the package insert. |
| Jul 18, 2016 | Class II | Z-2548-2016 | Xpert MRSA
Catalog GXMRSA-120, Rev. AJ
Microbiology:
The Cepheid Xpert MRSA assay… Internal investigation found the certain lots of MRSA kits have the potential for performance issues at lower positive sample… |
| Jul 27, 2015 | Class III | Z-2431-2015 | Xpert CT /NG Vaginal/Endocervical/ Specimen Collection Kit:
Part number… and 1 more product Firm determined that some lots of CT/NG Swab 50 and CT/NG Urine collection kits contain a sub component with an expiry date that is earlier… |
FDA record score
19 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
11 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0724-2026, this FEI as the recalling firm | Nov 5, 2025 | 4 x 1 | 4 |
| Class II recall Z-0959-2026, this FEI as the recalling firm | Oct 30, 2025 | 4 x 1 | 4 |
| Class II recall Z-0414-2026, this FEI as the recalling firm | Aug 6, 2025 | 4 x 1 | 4 |
| Class II recall Z-2584-2024, this FEI as the recalling firm | Jun 20, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0215-2024, this FEI as the recalling firm | Apr 24, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0592-2018, this FEI as the recalling firm | Oct 9, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0028-2018, this FEI as the recalling firm | Sep 15, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2548-2016, this FEI as the recalling firm | Jul 18, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 16, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 13, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 29, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing CEPHEID as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check CEPHEID yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Cepheid
- Location
- Sunnyvale, CA (US)
- Source
- device